INTRODUCTION

Advanced hybrid closed-loop (AHCL) systems have shown promise in improving glycemic control in adults with type 1 diabetes (T1D), yet real-world evidence remains limited. This study evaluated the impact of transitioning from multiple daily injections (MDI) or conventional continuous subcutaneous insulin infusion (CSII) to the Tandem t:slim X2™ Control-IQ® AHCL system on glycemic outcomes in adults with T1D.

METHODS

In this retrospective study, 56 non-pregnant adults with T1D were followed for 6 months. Primary outcomes were changes in the Glycemia Risk Index (GRI) and percentage time in the tight range (%TiTR70-140). Secondary outcomes included other standardized ambulatory glucose profile metrics as well as additional glycemic, anthropometric, and insulin dosing measures.

RESULTS

Transition to the Control-IQ system increased %TiTR70-140 by 11.5 percentage points and reduced GRI by 23.5 points. Both components of the GRI, hypoglycemia risk (%Chypo) and hyperglycemia risk (%Chyper), improved significantly. Hemoglobin A1c decreased by 1.0%, time in range (%TIR70-180) improved by 13.6 percentage points, and glycemic variability (CV%) showed marked improvement. Total daily insulin dose decreased by 34%, accompanied by modest weight loss. No diabetic ketoacidosis episodes were reported throughout the study period.

CONCLUSION

Transitioning to the Tandem t:slim X2 Control-IQ AHCL system led to significant improvements in tight glycemic control, glycemic risk reduction, and insulin efficiency, with a favorable safety profile. These findings support broader adoption of AHCL technology and underscore the need for prospective multicenter studies.

Для цитирования: Al Hayek Ayman, Alzahrani Wael M, Al Saeed Abdulghani H, Al Mashali Malak, Klonoff David C, Al Dawish Mohamed A. Evaluation of Glycemia Risk Index and Continuous Glucose Monitoring Outcomes Following the Transition to an Advanced Hybrid Closed-Loop System in Type 1 Diabetes.. Advances in therapy. 2025. DOI: 10.1007/s12325-025-03326-0.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.