MedElevator Войти в исследования
← Поиск публикаций

DOI 10.2106/jbjs.cc.24.00284

Медицинская библиотека · JBJS case connector · 2024

Total Hip Arthroplasty Broach Failure, Incarceration, and Retrieval: A Case Report.

Soderlund T, Ghulam-Jelani Z, Kollmorgen R

СтатьяКлинический случайПеревод в очереди

Аннотация

CASE: The direct anterior total hip arthroplasty is known for a steep learning curve and femoral-sided intraoperative complications. This is a case report of a failure of femoral broach with a subsequent incarcerated femoral broach and novel extraction technique during a direct anterior total hip arthroplasty. This rare complication has only one other report in the literature.

CONCLUSION

Femoral broach failure is a rare complication of direct anterior hip arthroplasty. Surgeons should be familiar with alternative extraction techniques to safely remove an incarcerated broach.

Для цитирования: Soderlund T, Ghulam-Jelani Z, Kollmorgen R. Total Hip Arthroplasty Broach Failure, Incarceration, and Retrieval: A Case Report.. JBJS case connector. 2024. DOI: 10.2106/jbjs.cc.24.00284.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.