Evaluation of pulse cooximetry in patients undergoing abdominal or pelvic surgery.

Applegate Richard L, Barr Steven J, Collier Carl E, Rook James L, Mangus Dustin B, Allard Martin W

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Intraoperative transfusion decisions generally are guided by blood loss estimation and periodic invasive hemoglobin measurement. Continuous hemoglobin measurement by pulse cooximetry (pulse hemoglobin; Rainbow® SET Pulse CO-Oximeter, Masimo Corporation, Irvine, CA) has good agreement with laboratory hemoglobin in healthy volunteers and could aid transfusion decision-making. Because intraoperative physiology may alter performance of this device, this study investigated pulse hemoglobin during surgery.

METHODS

Ninety-one adult patients undergoing abdominal or pelvic surgery in which large blood loss was likely were studied. Time-matched pulse hemoglobin measurements were recorded for each intraoperative arterial hemoglobin measurement obtained. Agreement between measurements was assessed by average difference (mean ± SD, g/dl), linear regression, and multiple measures Bland-Altman analysis.

RESULTS

The average difference between 360 time-matched measurements (bias) was 0.50 ± 1.44 g/dl, with wider limits of agreement (-2.3 to 3.3 g/dl) than reported in healthy volunteers. The average difference between 269 paired sequential pulse and arterial hemoglobin changes was 0.10 ± 1.11 g/dl, with half between -0.6 and 0.7 g/dl of each other. The bias was larger in patients with blood loss of more than 1,000 ml; hemoglobin less than 9.0 g/dl; any intraoperative transfusion; or intraoperative decrease in arterial hemoglobin at the time of sampling ≥2 g/dl (all P < 0.001). The range of bias was narrower at deeper anesthesia (P < 0.001).

CONCLUSIONS

Evaluation of the sensor and software version tested suggests that although pulse cooximetry may perform well in ambulatory subjects, in patients undergoing surgery in which large blood loss is likely, an invasive measurement should be used in transfusion decision-making.

Для цитирования: Applegate Richard L, Barr Steven J, Collier Carl E, Rook James L, Mangus Dustin B, Allard Martin W. Evaluation of pulse cooximetry in patients undergoing abdominal or pelvic surgery.. Anesthesiology. 2012. DOI: 10.1097/aln.0b013e31823d774f.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.