An amperometric horseradish peroxidase inhibition biosensor for the determination of phenylhydrazine.
Аннотация
An amperometric horseradish peroxidase (HRP) inhibition biosensor has been substantially constructed by the help of N,N-dicyclohexylcarbodiimide (DCC), N-hydroxysuccinimide (NHS). The preparation steps and the biosensor response to phenylhydrazine were monitored by electrochemical impedance spectroscopy (EIS), cyclic voltammetry, and chronoamperometry. The proposed biosensor could be applied to determine phenylhydrazine in a 0.10 M phosphate buffer solution containing 1.2 mM hydroquinone and 0.50 mM H(2)O(2) by phenylhydrazine, inhibiting the catalytic activity of the HRP enzyme in the reduction of H(2)O(2). The system was optimized to realize a reliable determination of phenylhydrazine in the range of 2.5 x 10(-7) to 1.1 x 10(-6) M with a detection limit of 8.2 x 10(-8) M and a correlation coefficient of 0.999. The modified electrode displayed good reproducibility, sensitivity and stability for the determination of phenylhydrazine.