A Multicenter Evaluation of a Nongel Mechanical Separator Plasma Blood Collection Tube for Testing of Selected Therapeutic Drugs.
СтатьяМногоцентровое исследованиеНаблюдательное исследованиеПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Morosyuk Svetlana, Berube Julie, Christenson Robert, Wu Alan H B, Uettwiller-Geiger Denise, Palicka Vladimir, Prusa Richard, Zima Tomas, Shaw Kristin, Plokhoy Elizabeth, Ahuja Aparna J, Stankovic Ana K
Аннотация
BACKGROUND: Some therapeutic drugs are unstable during sample storage in gel tubes. BD Vacutainer® Barricor™ Plasma Blood Collection Tube with nongel separator was compared with plasma gel tubes, BD Vacutainer PST™, PST II, and BD Vacutainer Serum Tube for acetaminophen, salicylate, digoxin, carbamazepine, phenytoin, valproic acid, and vancomycin during sample storage for up to 7 days.
METHODS: Seven hospital sites enrolled 705 participants who were taking at least one selected drug. The study tubes were collected and tested at initial time (0 h), after 48 h of storage at room temperature and on day 7 (after additional 5 days of refrigerated storage). The performance of BD Barricor tube was evaluated for each drug by comparing BD Barricor samples with samples from the other tubes at 0 h from the same participant; stability was evaluated by comparing test results from the same tube at 0 h, 48 h, and 7 days.
RESULTS: At 0 h, BD Barricor showed clinically equivalent results for selected therapeutic drugs compared with the other tubes, except phenytoin in BD PST. Phenytoin samples ≥20 µg/mL in BD PST had 10-12% lower values than samples in BD Barricor. During sample storage, all selected drugs remained stable for 7 days in BD Barricor and in serum aliquots. In BD PST, all drugs remained stable except phenytoin and carbamazepine and in BD PST II except for phenytoin.
CONCLUSION: The BD Barricor Tube is effective for the collection and storage of plasma blood samples for therapeutic drug monitoring without sample aliquoting.
METHODS: Seven hospital sites enrolled 705 participants who were taking at least one selected drug. The study tubes were collected and tested at initial time (0 h), after 48 h of storage at room temperature and on day 7 (after additional 5 days of refrigerated storage). The performance of BD Barricor tube was evaluated for each drug by comparing BD Barricor samples with samples from the other tubes at 0 h from the same participant; stability was evaluated by comparing test results from the same tube at 0 h, 48 h, and 7 days.
RESULTS: At 0 h, BD Barricor showed clinically equivalent results for selected therapeutic drugs compared with the other tubes, except phenytoin in BD PST. Phenytoin samples ≥20 µg/mL in BD PST had 10-12% lower values than samples in BD Barricor. During sample storage, all selected drugs remained stable for 7 days in BD Barricor and in serum aliquots. In BD PST, all drugs remained stable except phenytoin and carbamazepine and in BD PST II except for phenytoin.
CONCLUSION: The BD Barricor Tube is effective for the collection and storage of plasma blood samples for therapeutic drug monitoring without sample aliquoting.