DOI 10.1016/j.bios.2008.01.030
CYP450 biosensors based on gold chips for antiepileptic drugs determination.
Аннотация
Three-electrode configuration chips containing a Pt, Au and a screen-printed Ag/AgCl as counter, working and reference electrode, respectively, have been developed. Selective determination of Phenobarbital (PB) has been carried out by Cytochrome P450 2B4 (CYP450) immobilization into a polypyrrole matrix onto the gold working electrode. Chronoamperometric experiments show a PB diffusion coefficient of 2.42x10(-6)cm(2)s(-1), a reproducibility and repeatability in terms of residual standard deviation (RSD) of 13% and 5.51%, respectively, and a limit of detection (LOD) of 0.289 micromol dm(-3) (alpha=beta=0.05) for the developed CYP450-biosensor chip. Its performance has been showed by the determination of PB in pharmaceutical drugs. HPLC has been used as reference technique.