Comparison of the QMS Analyzer With HPLC-UV for the Quantification of Lamotrigine Concentrations in Human Plasma Samples.

Baldelli Sara, Castoldi Simone, Charbe Nitin, Cozzi Valeria, Fucile Serena, Cattaneo Dario, Clementi Emilio

Сравнительное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND

Recently, a turbidimetric immunoassay method has been developed for use in the form of a QMS lamotrigine (LTG) commercial immunoassay. This study was designed to evaluate the performance of this immunoassay using a validated high-performance liquid chromatography-ultraviolet (HPLC-UV) method as the reference.

METHODS

The performance of QMS was initially tested using drug-free plasma spiked with different amounts of LTG and, subsequently, by analyzing 61 trough plasma samples from epileptic patients given the drug as part of their maintenance antiepileptic therapies.

RESULTS

The correlation between LTG concentrations measured by QMS and HPLC was good, with a Pearson coefficient of 0.968 (P < 0.0001). The Bland-Altman approach showed that LTG concentrations measured with QMS exceeded HPLC on an average by 15.6% (limits of agreement, -18% to +63%), with a concentration-dependent performance (mean percent bias, 49.5 ± 8.2% and 0.6 ± 12.7% for concentrations less than 2 mg/L and greater than 14.9 mg/L, respectively).

CONCLUSIONS

The QMS provided acceptable analytical performance across a wide concentration range for routine LTG measurements, being at least comparable with the other commercial immunoassays. It could be, therefore, considered as a viable alternative to HPLC methods for routine LTG monitoring in the clinical practice, although its suitability for accurate analysis of samples with low concentration is limited.

Для цитирования: Baldelli Sara, Castoldi Simone, Charbe Nitin, Cozzi Valeria, Fucile Serena, Cattaneo Dario, Clementi Emilio. Comparison of the QMS Analyzer With HPLC-UV for the Quantification of Lamotrigine Concentrations in Human Plasma Samples.. Therapeutic drug monitoring. 2015. DOI: 10.1097/ftd.0000000000000202.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.