OBJECTIVE

Imageless computer assisted total knee surgical systems have commonly relied on determination of the functional rotational center of the femoral head as a landmark for determining the lower extremity mechanical axis. This has been accomplished through range of motion and center of rotation calculations for the femur with respect to the pelvis as the lower extremity is taken through a range of motion. Our study evaluated the use of this algorithm with and without a pelvic tracker attached to the iliac crest.

MATERIALS AND

METHODS

The functional center of the hip joint was also compared to the true radiographic center as determined by spiral CT data. Evaluating the different methods on six lower extremities from three whole-body cadavers revealed significant differences in the location of the calculated hip joint center, but little difference in the resulting lower extremity mechanical axis determination. The functional hip joint centers measured with and without a pelvic tracker differed from one another and from the CT-determined hip center.

RESULTS

No differences were found in the coronal plane measurements, but statistically significant differences were found in the sagittal plane measurements.

CONCLUSION

Algorithms that reduce the noise generated by pelvic movement should be devised to eliminate the need for a pelvic tracker.

Для цитирования: Mihalko William M, Phillips Matthew J, Fishkin Zair, Krackow Kenneth A. Pelvic tracker effects on hip center accuracy using imageless navigation.. Computer aided surgery : official journal of the International Society for Computer Aided Surgery. 2006. DOI: 10.3109/10929080600942147.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.