DOI 10.1016/j.ijpharm.2026.127389

Медицинская библиотека · International journal of pharmaceutics · 2026

Fluorinated radiopharmaceuticals sorption onto plasticized polyvinyl chloride automated injector tubings: implications for patient dosing.

Kazek Lola, Tylski Perrine, Flaus Anthime, Bolot Claire, Levigoureux Elise, Le Basle Yoann, Feschet-Chassot Emmanuelle, Mailhot-Jensen Bénédicte, Durand Alice, Chabanse Atekka, Chennell Philip, Chaib Sarah

СтатьяПеревод в очереди

Аннотация

BACKGROUND

The use of automated injectors for the administration of radiopharmaceuticals limits staff exposure and optimises the dose administered to patients, but many of these devices are only validated for a limited number of radiopharmaceuticals. The objectives of this study were to investigate the physicochemical phenomena occurring during the manipulation of fluorinated radiopharmaceuticals when in contact with injector tubings and to quantify and qualify the interactions between the radiopharmaceuticals and the tubing materials.

RESULTS

For six of the radiopharmaceuticals tested, the measured doses reached 97.64 ± 0.24% of the programmed dose. These results indicate an absence of sorption phenomena and validate their use on the injector. In contrast, the three most lipophilic molecules caused the injector to malfunction, resulting in significant residual activity in the tubing of up to 67%. These results were confirmed by activimeter measurements and positron emission tomography images.

RESULTS

from electronic microscopy results showed an alteration of the inner surface of the tubings that came into contact with lipophilic radiopharmaceuticals; however, this modification did not drastically alter the surface Zeta potential. Infrared spectroscopy confirmed the composition of the tubings as being plasticized polyvinyl chloride (PVCp), and plasticizer identification and quantification showed the presence of di-(2-ethylhexyl) adipate (reaching 37.5%) and di-(2-ethylhexyl) phthalate (more than 20%). A clear improvement was observed when co-extruded polyethylene / PVPp tubings were used.

CONCLUSIONS

Unlike lipophilic radiopharmaceuticals, which exhibit sorption behaviour dependent on the presence of plasticizers, the tested hydrophilic radiopharmaceuticals demonstrated minimal interaction with PVCp tubing and can be reliably administered using the automated injector.

Для цитирования: Kazek Lola, Tylski Perrine, Flaus Anthime, Bolot Claire, Levigoureux Elise, Le Basle Yoann, Feschet-Chassot Emmanuelle, Mailhot-Jensen Bénédicte, Durand Alice, Chabanse Atekka, Chennell Philip, Chaib Sarah. Fluorinated radiopharmaceuticals sorption onto plasticized polyvinyl chloride automated injector tubings: implications for patient dosing.. International journal of pharmaceutics. 2026. DOI: 10.1016/j.ijpharm.2026.127389.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.