MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1007/s11701-026-03751-5

Медицинская библиотека · Journal of robotic surgery · 2026

Comparative evaluation of CT-to-Fluoroscopy and Scan-and-Plan registration techniques in robot-assisted pedicle screw placement: a preliminary cadaveric faesibility study.

Oh Jacob, Ruangchainikom Monchai, Kripalani Satish, Vidyadhara S, Beng Saw Lim, Yap Wayne Ming Quan, Manghwani Jitesh, Dave Mirant, Chhabra H S

СтатьяСравнительное исследованиеПеревод в очереди

Аннотация

STUDY

DESIGN

Preliminary Cadaveric Faesibility study.

PURPOSE: To evaluate the preliminary faesibility, technical accuracy, and procedural metrics of CT-to-Fluoroscopy and Scan-and-Plan registration techniques in robot-assisted pedicle screw placement using the Mazor X robotic system.

BACKGROUND

Robot-assisted pedicle screw placement has been associated with improved accuracy and safety compared with conventional freehand techniques. However, direct comparisons between the two commonly used robotic registration workflows-CT-to-Fluoroscopy and Scan-and-Plan-remain limited. Evaluating their relative performance is important for optimizing surgical workflow and radiation utilization.

METHODOLOGY: Four fresh-frozen male cadavers were allocated to CT-to-Fluoroscopy (n = 2) or Scan-and-Plan (n = 2) registration groups. Eight experienced spine surgeons inserted a total of 32 pedicle screws bilaterally from L2 to L5. Screw accuracy was assessed using the Gertzbein-Robbins classification and trajectory deviation analysis in axial, sagittal, and coronal planes. Procedural time, surgeon-reported ease of use, and radiation exposure were also evaluated.

RESULTS

All 32 screws were clinically acceptable (Grade A + B), resulting in 100% overall accuracy in both groups. CT-to-Fluoroscopy demonstrated a higher proportion of Grade A screws than Scan-and-Plan (87.5% vs. 50.0%; p = 0.054). Scan-and-Plan showed significantly lower sagittal angular deviation (0.55 ± 0.71° vs. 1.66 ± 1.14°; p = 0.002), superior clearance (0.20 ± 0.16 mm vs. 0.45 ± 0.36 mm; p = 0.038), and lower coronal angular deviation (0.59 ± 0.76° vs. 1.99 ± 1.74°; p = 0.002). Planning, execution, and overall procedure times were comparable between groups, although Scan-and-Plan required shorter registration time. Surgeon satisfaction was identical (4.75 ± 0.46). Radiation exposure to surgeons was negligible in both groups, while cadaver radiation exposure was higher with CT-to-Fluoroscopy.

CONCLUSION

Both registration workflows achieved excellent accuracy. Scan-and-Plan demonstrated advantages in registration efficiency and patient-equivalent radiation exposure while maintaining comparable procedural performance, supporting its use when workflow optimization and radiation reduction are priorities.

Для цитирования: Oh Jacob, Ruangchainikom Monchai, Kripalani Satish, Vidyadhara S, Beng Saw Lim, Yap Wayne Ming Quan, Manghwani Jitesh, Dave Mirant, Chhabra H S. Comparative evaluation of CT-to-Fluoroscopy and Scan-and-Plan registration techniques in robot-assisted pedicle screw placement: a preliminary cadaveric faesibility study.. Journal of robotic surgery. 2026. DOI: 10.1007/s11701-026-03751-5.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.