m+Medelevator.RU
← Поиск публикаций
Investigative ophthalmology & visual science · 2016

Aqueous Flow Measured by Fluorophotometry in the Mouse.

СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Toris Carol B, Fan Shan, Johnson Thomas V, Camras Lucinda J, Hays Cassandra L, Liu Hong, Ishimoto Bruce M

Аннотация

PURPOSE: A fluorophotometer designed to measure aqueous flow in murine eyes was tested with artificial fluorescein chambers and in live mice with different anesthesia regimens, aqueous flow suppressants, and an anterior chamber cannulation method.

METHODS: Two hours following topical fluorescein application, one group of CD-1 mice was anesthetized with ketamine/xylazine, 2,2,2-tribromoethanol, or ketamine alone. Cornea and anterior chamber fluorescein concentrations were measured periodically for 60 to 90 minutes by fluorophotometric scans to calculate aqueous flow. Later, a subgroup of mice underwent aqueous flow measurement by anterior chamber cannulation. A third group was treated with timolol, dorzolamide, and vehicle in a crossover manner 1 hour prior to fluorophotometric scans.

RESULTS: Aqueous flow with ketamine/xylazine anesthesia (0.09 ± 0.05 μL/min, mean ± SD, n = 24) was slower than with tribromoethanol or ketamine alone (P < 0.001). Timolol reduced aqueous flow from 0.20 ± 0.07 μL/min to 0.07 ± 0.03 μL/min (P = 0.001) under tribromoethanol anesthesia and from 0.14 ± 0.03 μL/min to 0.10 ± 0.02 μL/min (P = 0.004) under ketamine anesthesia but not under ketamine/xylazine anesthesia. Dorzolamide reduced aqueous flow from 0.09 ± 0.03 to 0.06 ± 0.03 μL/min (P = 0.04) under ketamine/xylazine anesthesia. Aqueous flow by anterior chamber cannulation (0.20 ± 0.13 μL/min) was greater (P = 0.05) than by fluorophotometry (0.09 ± 0.07 μL/min).

CONCLUSIONS: A new noninvasive fluorophotometric method detected effects of general anesthesia and known aqueous suppressants on aqueous flow in mice. Aqueous flow measured by fluorophotometry was slower than by cannulation, and was technically easier with less variability. The mouse fluorophotometer is useful for repeated measurements of aqueous flow in the murine eye making crossover and longitudinal studies possible.

Предметные рубрики MeSH

ИП Урсолов Владлен Сергеевич

ИНН 500118482352 · ОГРНИП 314500117700049

Реквизиты и контакты