MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.gie.2018.01.025

Медицинская библиотека · Gastrointestinal endoscopy · 2018

Impact of 2 generational improvements in colonoscopes on adenoma miss rates: results of a prospective randomized multicenter tandem study.

Pioche Mathieu, Denis Angélique, Allescher Hans-Dieter, Andrisani Gianluca, Costamagna Guido, Dekker Evelien, Fockens Paul, Gerges Christian, Groth Stefan, Kandler Jennis, Lienhart Isabelle, Neuhaus Horst, Petruzziello Lucio, Schachschal Guido, Tytgat Kristien, Wallner Jürgen, Weingart Vincens, Touzet Sandrine, Ponchon Thierry, Rösch Thomas

СтатьяМногоцентровое исследованиеРандомизированное исследованиеПеревод в очереди

Аннотация

BACKGROUND AND AIMS: Numerous randomized studies have shown that changing certain features of colonoscopes, usually incorporated when switching from one endoscope generation to the next, mostly do not increase adenoma yield. There is, however, indirect evidence that it may be necessary to skip one instrument generation (ie, changing from one generation to the next but one) to achieve this effect.

METHODS

We compared the latest-generation colonoscopes from one company (Olympus Exera III, 190-C) with the next to last one (Olympus 160/5-C) in a prospective multicenter study randomized for the order of colonoscopes in a tandem fashion, involving 2 different examiners. Patients with increased risk for colorectal neoplasia undergoing colonoscopy (positive fecal occult blood test, personal/familial history of colorectal cancer/adenoma, rectal bleeding, recent change in bowel movements) were included. The primary outcome was the adenoma miss rate with the 190 (190-C) colonoscope in comparison with the 160/5 colonoscope (160/5-C).

RESULTS

A total of 856 patients (48.8% male; mean age, 58.3 years) with a personal (41%) or family (38%) history of colorectal neoplasia, rectal bleeding (19%), and other indications were included. Of the 429 patients in the 190-C first group, 16.6% (95% confidence interval [CI], 13.0%-20.1%) had at least one adenoma missed during the first procedure, compared with 30.2% (95% CI, 25.9%-34.6%) in the group with 160/5-C first (P < .001). Similarly, the adenoma detection rate during the first colonoscopy was 43.8% versus 36.5% (P = .030) for 190-C versus 160/5-C, respectively.

CONCLUSIONS

This randomized tandem trial showed lower adenoma miss rates and higher adenoma detection rates for the newer 190 colonoscopes compared with the 160/5 series. These results suggest that it takes multiple improvements, such as those implemented over 2 instrument generations, before an effect on adenoma (miss) rate can be observed. (Study registration number: ISRCTN 2010-A01256-33.).

Для цитирования: Pioche Mathieu, Denis Angélique, Allescher Hans-Dieter, Andrisani Gianluca, Costamagna Guido, Dekker Evelien, Fockens Paul, Gerges Christian, Groth Stefan, Kandler Jennis, Lienhart Isabelle, Neuhaus Horst, Petruzziello Lucio, Schachschal Guido, Tytgat Kristien, Wallner Jürgen, Weingart Vincens, Touzet Sandrine, Ponchon Thierry, Rösch Thomas. Impact of 2 generational improvements in colonoscopes on adenoma miss rates: results of a prospective randomized multicenter tandem study.. Gastrointestinal endoscopy. 2018. DOI: 10.1016/j.gie.2018.01.025.