Strut penetration: local complications, breakthrough pulmonary embolism, and retrieval failure in patients with Celect vena cava filters.
СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Bos Aaron, Van Ha Thuong, van Beek Darren, Ginsburg Michael, Zangan Steven, Navuluri Rakesh, Lorenz Jonathan, Funaki Brian
Аннотация
PURPOSE: To investigate strut penetration in patients with Celect filters, specifically local complications and association with breakthrough pulmonary embolism (PE) or retrieval failure.
MATERIALS AND METHODS: A retrospective single-center study was conducted to evaluate patients who received Celect filters between January 2007 and May 2013. A total of 595 filters were placed during the study period. Primary indications included thromboembolic disease (93%) and primary surgical prophylaxis (7%). Complications and retrieval data were assessed by computed tomography (CT) and electronic medical records.
RESULTS: A total of 193 patients underwent follow-up abdominal CT at a mean follow-up interval of 176.2 days (range, 0-1,739 d). The rate of strut penetration more than 3 mm outside the caval wall was 28.5% (n = 55). One patient had CT evidence of clinically major strut penetration (1.8%) with strut compression of the right ureter causing hydronephrosis. Indwelling filter time longer than 100 days was associated with strut penetration (P < .001). Age, sex, and history of thromboembolic disease were not associated with strut penetration (P = .51, P = .81, and P = .89). Sixty-three patients presented for follow-up CT pulmonary angiography at a mean of 128.1 days (range, 1-895 d). The rate of breakthrough PE was 12.7%. The overall retrieval success rate was 96.7% (n = 150). Strut penetration was not associated with breakthrough PE or retrieval failure (P = .49 and P = .22).
CONCLUSIONS: Although strut penetration is a common complication with Celect filters, there is no association with breakthrough PE or retrieval failure. CT evidence of local complications associated with strut penetration is rare.
MATERIALS AND METHODS: A retrospective single-center study was conducted to evaluate patients who received Celect filters between January 2007 and May 2013. A total of 595 filters were placed during the study period. Primary indications included thromboembolic disease (93%) and primary surgical prophylaxis (7%). Complications and retrieval data were assessed by computed tomography (CT) and electronic medical records.
RESULTS: A total of 193 patients underwent follow-up abdominal CT at a mean follow-up interval of 176.2 days (range, 0-1,739 d). The rate of strut penetration more than 3 mm outside the caval wall was 28.5% (n = 55). One patient had CT evidence of clinically major strut penetration (1.8%) with strut compression of the right ureter causing hydronephrosis. Indwelling filter time longer than 100 days was associated with strut penetration (P < .001). Age, sex, and history of thromboembolic disease were not associated with strut penetration (P = .51, P = .81, and P = .89). Sixty-three patients presented for follow-up CT pulmonary angiography at a mean of 128.1 days (range, 1-895 d). The rate of breakthrough PE was 12.7%. The overall retrieval success rate was 96.7% (n = 150). Strut penetration was not associated with breakthrough PE or retrieval failure (P = .49 and P = .22).
CONCLUSIONS: Although strut penetration is a common complication with Celect filters, there is no association with breakthrough PE or retrieval failure. CT evidence of local complications associated with strut penetration is rare.