Strut penetration: local complications, breakthrough pulmonary embolism, and retrieval failure in patients with Celect vena cava filters.

Bos Aaron, Van Ha Thuong, van Beek Darren, Ginsburg Michael, Zangan Steven, Navuluri Rakesh, Lorenz Jonathan, Funaki Brian

СтатьяПеревод в очереди

Аннотация

PURPOSE: To investigate strut penetration in patients with Celect filters, specifically local complications and association with breakthrough pulmonary embolism (PE) or retrieval failure.

MATERIALS AND

METHODS

A retrospective single-center study was conducted to evaluate patients who received Celect filters between January 2007 and May 2013. A total of 595 filters were placed during the study period. Primary indications included thromboembolic disease (93%) and primary surgical prophylaxis (7%). Complications and retrieval data were assessed by computed tomography (CT) and electronic medical records.

RESULTS

A total of 193 patients underwent follow-up abdominal CT at a mean follow-up interval of 176.2 days (range, 0-1,739 d). The rate of strut penetration more than 3 mm outside the caval wall was 28.5% (n = 55). One patient had CT evidence of clinically major strut penetration (1.8%) with strut compression of the right ureter causing hydronephrosis. Indwelling filter time longer than 100 days was associated with strut penetration (P < .001). Age, sex, and history of thromboembolic disease were not associated with strut penetration (P = .51, P = .81, and P = .89). Sixty-three patients presented for follow-up CT pulmonary angiography at a mean of 128.1 days (range, 1-895 d). The rate of breakthrough PE was 12.7%. The overall retrieval success rate was 96.7% (n = 150). Strut penetration was not associated with breakthrough PE or retrieval failure (P = .49 and P = .22).

CONCLUSIONS

Although strut penetration is a common complication with Celect filters, there is no association with breakthrough PE or retrieval failure. CT evidence of local complications associated with strut penetration is rare.

Для цитирования: Bos Aaron, Van Ha Thuong, van Beek Darren, Ginsburg Michael, Zangan Steven, Navuluri Rakesh, Lorenz Jonathan, Funaki Brian. Strut penetration: local complications, breakthrough pulmonary embolism, and retrieval failure in patients with Celect vena cava filters.. Journal of vascular and interventional radiology : JVIR. 2014. DOI: 10.1016/j.jvir.2014.09.010.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.