MedElevator Войти в исследования
← Поиск публикаций

DOI 10.5301/ijao.5000225

Медицинская библиотека · The International journal of artificial organs · 2013

Extracorporeal membrane oxygenation cannula malposition in the azygos vein in an infant.

Balmaks Reinis, Sufficool Kari E, White Frances V, Byrge Melissa M, Kolovos Nikoleta S

Клинический случайСтатьяПеревод в очереди

Аннотация

INTRODUCTION

Misplacement of extracorporeal membrane oxygenation (ECMO) venous cannula in the azygos vein has previously been described only in newborns.

CASE: For the first time, we report an aberrant ECMO cannula placement in the azygos vein in a child outside neonatal period. History of chronic lung disease and ligation of persistent ductus arteriosus leading to elevated right ventricular pressure was the potential risk factor.

CONCLUSIONS

This rare complication should be considered whenever the patient has inadequate venous return on ECMO and it can be easily ruled out with a lateral chest radiograph.

Для цитирования: Balmaks Reinis, Sufficool Kari E, White Frances V, Byrge Melissa M, Kolovos Nikoleta S. Extracorporeal membrane oxygenation cannula malposition in the azygos vein in an infant.. The International journal of artificial organs. 2013. DOI: 10.5301/ijao.5000225.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.