Combined saturation/inversion recovery sequences for improved evaluation of scar and diffuse fibrosis in patients with arrhythmia or heart rate variability.
Аннотация
PURPOSE: To develop arrhythmia-insensitive inversion recovery sequences for improved visualization of myocardial scar and quantification of diffuse fibrosis.
METHODS
A novel preparation pre-pulse, called saturation pulse prepared heart-rate-independent inversion recovery, is introduced, which consists of a combination of saturation and inversion pulses to remove the magnetization history in each heartbeat in late gadolinium enhancement (LGE) imaging and eliminate the need for rest periods in T1 mapping. The proposed LGE and T1 mapping sequences were evaluated against conventional LGE and modified Look-Locker inversion sequences using numerical simulations, phantom and imaging in healthy subjects and patients with suspected or known cardiovascular disease.
RESULTS
Simulations and phantom experiments show that the saturation pulse prepared heart-rate-independent inversion recovery pre-pulse in LGE reduces ghosting artifacts and results in perfect nulling of the healthy myocardium in the presence of arrhythmia. In T1 mapping, saturation pulse prepared heart-rate-independent inversion recovery results in (a) reduced scan time (17 vs. 9 heartbeats), (b) insensitivity to heart rate for long T1, and (c) increased signal homogeneity for short T1. LGE images in a patient in atrial fibrillation during the scan show improved myocardial nulling. In vivo T1 maps demonstrate increased signal homogeneity in blood pools and myocardium.
CONCLUSION
The proposed sequences are insensitive to heart rate variability, yield improved LGE images in the presence of arrhythmias, as well as T1 mapping with shorter scan times.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.