An Open-Label Study Evaluating the Performance of the Dose Indicator in a Metered Dose Inhaler Delivering Glycopyrrolate and Formoterol Fumarate in Patients with Moderate-to-Very Severe Chronic Obstructive Pulmonary Disease.
Клиническое исследованиеСтатьяМногоцентровое исследованиеПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Pudi Krishna, Feldman Gregory, Fakih Faisal, Mack Peter, Maes Andrea, Siddiqui Shahid, St Rose Earl, Reisner Colin
Аннотация
BACKGROUND: GFF MDI is a glycopyrrolate/formoterol fumarate fixed-dose combination metered dose inhaler formulated using co-suspension delivery technology. This open-label, single-arm multicenter study (NCT02268396) evaluated the accuracy, reliability, and functionality of the GFF MDI AeroCount® dose indicator when used by patients with chronic obstructive pulmonary disease (COPD).
PATIENTS AND METHODS: The study enrolled subjects (40-80 years of age) with an established clinical history (≥6 months) of COPD, who completed an electronic diary twice daily to record study-drug administration time, the number of actuations used, and pre- and post-dose dose indicator readings. The primary endpoint was the percentage of devices for which the number of subject-reported actuations was consistent (±20 actuations) with the dose indicator-based actuation count (equal to 130 minus the dose indicator reading) at the end of the treatment period (4 weeks). Safety was monitored throughout the study.
RESULTS: A total of 138 subjects with moderate-to-very severe COPD (50.7% male; mean [standard deviation (SD)] age 62.1 [8.3] years) were enrolled and treated. Subject-reported actuation count and dose indicator-based actuation counts were consistent for 96.4% (132/137) of devices at the end of the treatment period (4 weeks) in the intent-to-treat (ITT) population and for all devices in the per-protocol (PP) population. The mean (SD) dose indicator-based actuation and subject-reported actuation counts in the ITT population (n = 137) were 113.4 (18.9) and 117.0 (19.0), respectively, with a mean (SD) difference of 3.6 (7.9). The mean (SD) dose indicator-based actuation and subject-reported actuation counts in the PP population (n = 112) were 116.8 (8.7) and 119.7 (8.1), respectively. There were no unexpected safety findings.
CONCLUSIONS: This study supported the accuracy, reliability, and utility of the dose indicator integrated into the GFF MDI device when used by patients with COPD.
PATIENTS AND METHODS: The study enrolled subjects (40-80 years of age) with an established clinical history (≥6 months) of COPD, who completed an electronic diary twice daily to record study-drug administration time, the number of actuations used, and pre- and post-dose dose indicator readings. The primary endpoint was the percentage of devices for which the number of subject-reported actuations was consistent (±20 actuations) with the dose indicator-based actuation count (equal to 130 minus the dose indicator reading) at the end of the treatment period (4 weeks). Safety was monitored throughout the study.
RESULTS: A total of 138 subjects with moderate-to-very severe COPD (50.7% male; mean [standard deviation (SD)] age 62.1 [8.3] years) were enrolled and treated. Subject-reported actuation count and dose indicator-based actuation counts were consistent for 96.4% (132/137) of devices at the end of the treatment period (4 weeks) in the intent-to-treat (ITT) population and for all devices in the per-protocol (PP) population. The mean (SD) dose indicator-based actuation and subject-reported actuation counts in the ITT population (n = 137) were 113.4 (18.9) and 117.0 (19.0), respectively, with a mean (SD) difference of 3.6 (7.9). The mean (SD) dose indicator-based actuation and subject-reported actuation counts in the PP population (n = 112) were 116.8 (8.7) and 119.7 (8.1), respectively. There were no unexpected safety findings.
CONCLUSIONS: This study supported the accuracy, reliability, and utility of the dose indicator integrated into the GFF MDI device when used by patients with COPD.
Предметные рубрики MeSH
Administration, InhalationAdrenergic beta-2 Receptor AgonistsAdultAgedAged, 80 and overBronchodilator AgentsDrug CombinationsEquipment DesignFemaleFormoterol FumarateGlycopyrrolateHumansLungMaleMetered Dose InhalersMiddle AgedMuscarinic AntagonistsPulmonary Disease, Chronic ObstructiveSeverity of Illness IndexTreatment OutcomeUnited States