MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1002/lary.27569

Медицинская библиотека · The Laryngoscope · 2018

How design characteristics of tracheostomy tubes affect the cannula and tracheal flows.

Subramaniam Dhananjay Radhakrishnan, Willging J Paul, Gutmark Ephraim J, Oren Liran

Оценочное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

OBJECTIVES

The aim of this study was to perform computational simulations of airflow within an anatomically accurate model of an adult trachea in different tracheostomy tube designs. We hypothesized that tracheal airflow in patients is significantly influenced by the geometry and size of these devices.

METHODS

The three-dimensional (3D) geometry of the trachea was reconstructed using computed tomography scans for an adult with no history of lung disease. 3D models of four cuffed tube designs, namely Tracoe, Portex, and Shiley Proximal and Distal tracheostomy tubes were generated using geometric modeling software. Transient simulations of airflow in the tube-airway assembly were performed for each tube using computational fluid dynamics (CFD).

RESULTS

Airflow velocity was higher for the Shiley tubes compared with Portex and Tracoe tubes. For all designs, the largest magnitude of inspiratory airflow turbulence was obtained midway in the trachea. The work of breathing, quantified by the resistance of the tracheostomy tube, was lowest for Tracoe. Maximum airway wall shear stress (WSS), defined as flow-induced frictional forces, occurred at the same spatial location in all cases. Low inspiratory WSS at the carina and high expiratory airway WSS at the cuff-airway interface were observed for the Tracoe and Portex tubes.

CONCLUSION

Our CFD model offers a promising approach not only for choosing a tracheostomy tube for a patient but for improving existing tracheostomy tube designs.

LEVEL OF EVIDENCE: NA Laryngoscope, 129:1791-1799, 2019.

Для цитирования: Subramaniam Dhananjay Radhakrishnan, Willging J Paul, Gutmark Ephraim J, Oren Liran. How design characteristics of tracheostomy tubes affect the cannula and tracheal flows.. The Laryngoscope. 2018. DOI: 10.1002/lary.27569.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.