Maximum cardiac output during incremental exercise by first-pass radionuclide ventriculography.

Maroni J M, Oelberg D A, Pappagianopoulos P, Boucher C A, Systrom D M

Сравнительное исследованиеСтатьяПоддержка исследования: вне правительства СШАПоддержка службы здравоохранения СШАПеревод в очереди

Аннотация

STUDY

OBJECTIVE

To validate a noninvasive first-pass radionuclide ventriculographic (FPRV) measurement of maximum cardiac output (Qv) during exercise.

DESIGN

Comparison of Qv to that measured by the Fick principle (Qf) at peak exercise.

SETTING

Academic cardiopulmonary exercise laboratory.

PATIENTS: Seventy-eight consecutive patients without a history of septal defect undergoing clinically indicated maximum incremental cardiopulmonary exercise testing with pulmonary arterial catheterization and FPRV.

MEASUREMENTS AND

RESULTS

Ventilation and gas exchange were measured breath-by-breath or by a mixing chamber/mass spectrometer system. Arterial and mixed venous O2 content were measured each minute during exercise. When patients without left-to-right ventricular stroke count ratio evidence for left-sided regurgitation were isolated, peak Qv was linearly related to Qf (r=0.75, p=0.0001). To account for a small systematic overestimation (bias) of Qf by Qv, the linear equation for the Qv/Qf relation was derived for patients studied between 1990 and 1993 and applied to those studied subsequently. The resulting corrected peak Qv was tightly related to peak Qf (r=0.90, p<0.001) with confidence intervals for slope and intercept overlapping identity.

CONCLUSION

FPRV can reasonably estimate maximum cardiac output during incremental exercise in patients for whom the technique has ruled out left-sided cardiac regurgitant lesions.

Для цитирования: Maroni J M, Oelberg D A, Pappagianopoulos P, Boucher C A, Systrom D M. Maximum cardiac output during incremental exercise by first-pass radionuclide ventriculography.. Chest. 1998. DOI: 10.1378/chest.114.2.457.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.