DOI 10.1515/cclm-2025-1499

Медицинская библиотека · Clinical chemistry and laboratory medicine · 2026

Laboratory solution to diagnose and monitor atypical IgG4-mediated anti-GBM disease.

Jacobs Joannes F M, Lambeck Annechien J A, Lammerts Rosa G M, van de Logt Anne-Els, Sanders Jan-Stephan, Lenssen Mark, Eijgelsheim Mark, Bajema Ingeborg, Steenbergen Eric J, Hoefsloot Wibe, van der Meijden Wilbert A G, van der Molen Renate G

СтатьяКлинический случайПеревод в очереди

Аннотация

OBJECTIVES

Diagnosing atypical IgG4-mediated anti-glomerular basement membrane (anti-GBM) disease is challenging because conventional serological assays poorly detect IgG4 antibodies. Here we study which commercial assays are affected and we describe proof-of-concept of improved IgG4 anti-GBM detection.

METHODS

To investigate the scope of the diagnostic dilemma detecting IgG4 anti-GBM antibodies, serum from a patient with atypical IgG4-mediated anti-GBM was distributed to 36 laboratories participating in the Dutch External Quality Assessment (EQA) program. To improve IgG4 anti-GBM detection in the fluorescent enzyme immunoassay (FEIA), the standard anti-IgG conjugate was replaced with anti-IgG4 conjugate.

RESULTS

We report the diagnostic delay of a patient with atypical anti-GBM disease who presented with an indolent disease course. Histopathology comprised hallmarks of classic anti-GBM disease including bright linear IgG deposits along the GBM but without the typical findings of diffuse crescentic and necrotizing glomerulonephritis. IgG anti-GBM test results were repeatedly negative. Histopathological subclass analysis demonstrated that the linear IgG deposits were predominantly IgG4. All 36 Dutch EQA-participants reported negative anti-GBM test results, demonstrating that the diagnostic challenge of detecting IgG4 anti-GBM is broadly applicable across multiple diagnostic assays. A minor modification to the manufacturer's standard protocol of the FEIA anti-GBM dramatically improved the assay's performance for measuring antibodies of the IgG4 isotype. Using the modified IgG4-GBM test, we were able to diagnose and monitor one more patient with atypical IgG4-mediated anti-GBM disease.

CONCLUSIONS

Here we demonstrate proof-of-concept of a modified IgG4 anti-GBM blood test allowing serological confirmation of atypical anti-GBM disease and sensitive monitoring of therapy response.

Для цитирования: Jacobs Joannes F M, Lambeck Annechien J A, Lammerts Rosa G M, van de Logt Anne-Els, Sanders Jan-Stephan, Lenssen Mark, Eijgelsheim Mark, Bajema Ingeborg, Steenbergen Eric J, Hoefsloot Wibe, van der Meijden Wilbert A G, van der Molen Renate G. Laboratory solution to diagnose and monitor atypical IgG4-mediated anti-GBM disease.. Clinical chemistry and laboratory medicine. 2026. DOI: 10.1515/cclm-2025-1499.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.