Journal of biomedical optics · 2025

Anthropomorphic tissue-mimicking phantoms for oximetry validation in multispectral optical imaging.

СтатьяПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Dreher Kris K, Gröhl Janek, Grace Friso, Ayala Leonardo, Nölke Jan-Hinrich, Bender Christoph J, Watt Melissa J, White Katie-Lou, Tao Ran, Johnen Wibke, Tizabi Minu D, Seitel Alexander, Maier-Hein Lena, Bohndiek Sarah E

Аннотация

SIGNIFICANCE: Optical imaging of blood oxygenation ( sO 2 ) can be achieved based on the differential absorption spectra of oxy- and deoxyhemoglobin. A key challenge in realizing clinical validation of the sO 2 biomarkers is the absence of reliable sO 2 reference standards, including test objects.

AIM: To enable quantitative testing of multispectral imaging methods for assessment of sO 2 by introducing anthropomorphic phantoms with appropriate tissue-mimicking optical properties.

APPROACH: We used the stable copolymer-in-oil base material to create physical anthropomorphic structures and optimized dyes to mimic the optical absorption of blood across a wide spectral range. Using 3D-printed phantom molds generated from a magnetic resonance image of a human forearm, we molded the material into an anthropomorphic shape. Using both reflectance hyperspectral imaging (HSI) and photoacoustic tomography (PAT), we acquired images of the forearm phantoms and evaluated the performance of linear spectral unmixing (LSU).

RESULTS: Based on 10 fabricated forearm phantoms with vessel-like structures featuring five distinct sO 2 levels (between 0 and 100%), we showed that the measured absorption spectra of the material correlated well with HSI and PAT data with a Pearson correlation coefficient consistently above 0.8. Further, the application of LSU enabled a quantification of the mean absolute error in sO 2 assessment with HSI and PAT.

CONCLUSIONS: Our anthropomorphic tissue-mimicking phantoms hold potential to provide a robust tool for developing, standardising, and validating optical imaging of sO 2 .

Предметные рубрики MeSH