MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1002/mrm.70123

Медицинская библиотека · Magnetic resonance in medicine · 2025

Impact of Lower Limb Amputation on Implant Heating During 1.5 T MRI: Assessment of Active and Passive Device.

Shen Ao, Akter Mir Khadiza, Zuo Ziyu, Zhang Lingfei, Zheng Jianfeng, Steckner Michael, Du Mingtao, Chen Ji

СтатьяПеревод в очереди

Аннотация

PURPOSE: This study investigates the impact of lower limb amputation on radiofrequency (RF)-induced heating of active and passive implantable medical devices (AIMDs and PIMDs) in a 1.5 T MRI environment.

METHODS

High-resolution anatomical models representing various body types and levels of amputation (below-knee and above-knee) were explored in the study. Full-wave electromagnetic simulations were used to assess specific absorption rate (SAR) near modular hip prostheses. The RF-induced lead-tip heating for a clinically relevant peripheral nerve stimulation (PNS) device was evaluated following ISO 10974. Results were normalized using whole-body SAR and averaged B1+rms. Experimental validation was performed using an anatomically inspired leg phantom with embedded metallic implants.

RESULTS

SAR near passive implants increased with implant length and was influenced by the level of amputation. In general, more extensive amputations resulted in reduced SAR due to alterations in tissue geometry. For active implantable medical devices (AIMDs), temperature rise also decreased with greater amputation extent; however, localized heating effects were intensified near the amputation site. Phantom simulation results were in strong agreement with experimental measurements, supporting the accuracy of the modeling approach.

CONCLUSION

Limb amputation significantly alters RF field distribution, affecting RF-induced heating near both active and passive implants. These results highlight the need for amputation-specific safety assessments during MRI to ensure accurate evaluation of heating risks and implant compatibility in amputee patients.

Для цитирования: Shen Ao, Akter Mir Khadiza, Zuo Ziyu, Zhang Lingfei, Zheng Jianfeng, Steckner Michael, Du Mingtao, Chen Ji. Impact of Lower Limb Amputation on Implant Heating During 1.5 T MRI: Assessment of Active and Passive Device.. Magnetic resonance in medicine. 2025. DOI: 10.1002/mrm.70123.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.