Better precision of a new robotically assisted system for total knee arthroplasty compared to conventional techniques: A sawbone model study.

Li Zheng, Chen Xin, Zhang Xiaofeng, Yan Jun, Song Youdong, Huo Yujia, Lin Jin

СтатьяПеревод в очереди

Аннотация

BACKGROUND

The purpose of this study was to compare the accuracy of this new HURWA robotic-assisted total knee arthroplasty (TKA) technique to the accuracy of the conventional technique in a sawbone model.

METHODS

The HURWA robotic-assisted TKA system was applied in the robotic group. After bone resection, all of these sawbones were scanned by the use of a structured light scanning system. Measurements of bone resections, femoral coronal and sagittal measurements, and tibial coronal and sagittal measurements were recorded.

RESULTS

Compared to the conventional technique, the HURWA robotic-assisted system significantly improved the accuracy of the bone resection levels and angles. In the robotic group, the accuracy of all of the bone resection levels was below 0.6 mm (with standard deviation [SD] below 0.6 mm), and all of the bone resection angles were below 0.6° (with SD below 0.4°).

CONCLUSION

Our data suggest that this novel HURWA robotic-assisted system can significantly improve the accuracy of bone resection levels and angles.

Для цитирования: Li Zheng, Chen Xin, Zhang Xiaofeng, Yan Jun, Song Youdong, Huo Yujia, Lin Jin. Better precision of a new robotically assisted system for total knee arthroplasty compared to conventional techniques: A sawbone model study.. The international journal of medical robotics + computer assisted surgery : MRCAS. 2021. DOI: 10.1002/rcs.2263.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.