Endovascular foreign body removal of a MynxGrip polyethylene glycol (PEG) sealant embolus using a cerebral stent retriever.

Shamseldin Mohammed, Bergert Hendrik, Neumeister Axel, Puls Ralf

СтатьяПеревод в очереди

Аннотация

BACKGROUND

This is a rare case of removing an intra-arterial foreign body represented by MynxGrip polyethylene glycol (PEG) sealant as a rare complication of using the MynxGrip™ Vascular Closure Device (AccessClosure, Inc., Mountain View, CA) using a pRESET stent retriever (Phenox, Bochum, Germany) which is utilized mainly for treatment of endovascular stroke.

CASE PRESENTATION: A 60-year-old female patient suffering from intermittent claudication in the right lower limb (stage IIb according to Fontaine) due to a peripheral arterial occlusive disease was presented for an elective revascularization using balloon angioplasty of a short chronic occlusion of the right superficial femoral artery. After a successful revascularization of the right superficial femoral artery using a retrograde femoral access from the left common femoral artery, the patient suffered from an acute limb ischemia in the left foot with distal popliteal embolization with involvement of BTK (below the knee) trifurcation. This is believed to be due to an intra-arterial foreign body embolism of MynxGrip polyethylene glycol sealant as a rare complication of using the MynxGrip™ Vascular Closure Device.

CONCLUSIONS

Stent retrievers have been used previously in removing dislocated coils especially in the cerebral vessels. This case report however proves a high efficacy and safety of using stent retrievers in removing different and rather unusual intra-arterial foreign bodies such as MynxGrip polyethylene glycol sealant.

Для цитирования: Shamseldin Mohammed, Bergert Hendrik, Neumeister Axel, Puls Ralf. Endovascular foreign body removal of a MynxGrip polyethylene glycol (PEG) sealant embolus using a cerebral stent retriever.. CVIR endovascular. 2021. DOI: 10.1186/s42155-020-00197-0.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.