An open 8-channel parallel transmission coil for static and dynamic 7T MRI of the knee and ankle joints at multiple postures.

Jin Jin, Weber Ewald, Destruel Aurelien, O'Brien Kieran, Henin Bassem, Engstrom Craig, Crozier Stuart

СтатьяПеревод в очереди

Аннотация

PURPOSE: We present the initial in vivo imaging results of an open architecture eight-channel parallel transmission (pTx) transceive radiofrequency (RF) coil array that was designed and constructed for static and dynamic 7T MRI of the knee and ankle joints.

METHODS

The pTx coil has a U-shaped dual-row configuration (200 mm overall length longitudinally) that allows static and dynamic imaging of the knee and ankle joints at various postures and during active movements. This coil structure, in combination with B1 shimming, allows flexible configuration of B1 transmit profiles, with good homogeneity over 120-mm regions of interest. This coil enabled high-resolution gradient echo (e.g., 3D dual-echo steady state [DESS] and 3D multiecho data image combination [MEDIC]) and turbo spin echo (TSE) imaging (e.g., with proton density weighting [PDw], PDw with fat saturation, and T1 and T2 weightings) with local RF energy absorption rates well below regulatory limits.

RESULTS

High-resolution 2D and 3D image series (e.g., 0.3 mm in-plane resolution for TSE, 0.47 mm isotropic for DESS and MEDIC) were obtained from the knee and ankle joints with excellent tissue contrast. Dynamic imaging during continuous knee and ankle flexion-extension cycles were successfully acquired.

CONCLUSION

The new open pTx coil array provides versatility for high-quality static and dynamic MRI of the knee and ankle joints at 7T. Magn Reson Med 79:1804-1816, 2018. © 2017 International Society for Magnetic Resonance in Medicine.

Для цитирования: Jin Jin, Weber Ewald, Destruel Aurelien, O'Brien Kieran, Henin Bassem, Engstrom Craig, Crozier Stuart. An open 8-channel parallel transmission coil for static and dynamic 7T MRI of the knee and ankle joints at multiple postures.. Magnetic resonance in medicine. 2017. DOI: 10.1002/mrm.26804.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.