MedElevator Войти в исследования
← Поиск публикаций

DOI 10.11138/jts/2016.4.2.121

Медицинская библиотека · Joints · 2016

Evaluation of RSA set-up from a clinical biplane fluoroscopy system for 3D joint kinematic analysis.

Bonanzinga Tommaso, Signorelli Cecilia, Bontempi Marco, Russo Alessandro, Zaffagnini Stefano, Marcacci Maurilio, Bragonzoni Laura

СтатьяПеревод в очереди

Аннотация

PURPOSE: dinamic roentgen stereophotogrammetric analysis (RSA), a technique currently based only on customized radiographic equipment, has been shown to be a very accurate method for detecting three-dimensional (3D) joint motion. The aim of the present work was to evaluate the applicability of an innovative RSA set-up for in vivo knee kinematic analysis, using a biplane fluoroscopic image system. To this end, the Authors describe the set-up as well as a possible protocol for clinical knee joint evaluation. The accuracy of the kinematic measurements is assessed.

METHODS

the Authors evaluated the accuracy of 3D kinematic analysis of the knee in a new RSA set-up, based on a commercial biplane fluoroscopy system integrated into the clinical environment. The study was organized in three main phases: an in vitro test under static conditions, an in vitro test under dynamic conditions reproducing a flexion-extension range of motion (ROM), and an in vivo analysis of the flexion-extension ROM. For each test, the following were calculated, as an indication of the tracking accuracy: mean, minimum, maximum values and standard deviation of the error of rigid body fitting.

RESULTS

in terms of rigid body fitting, in vivo test errors were found to be 0.10±0.05 mm. Phantom tests in static and kinematic conditions showed precision levels, for translations and rotations, of below 0.1 mm/0.2° and below 0.5 mm/0.3° respectively for all directions.

CONCLUSIONS

the results of this study suggest that kinematic RSA can be successfully performed using a standard clinical biplane fluoroscopy system for the acquisition of slow movements of the lower limb.

CLINICAL RELEVANCE: a kinematic RSA set-up using a clinical biplane fluoroscopy system is potentially applicable and provides a useful method for obtaining better characterization of joint biomechanics.

Для цитирования: Bonanzinga Tommaso, Signorelli Cecilia, Bontempi Marco, Russo Alessandro, Zaffagnini Stefano, Marcacci Maurilio, Bragonzoni Laura. Evaluation of RSA set-up from a clinical biplane fluoroscopy system for 3D joint kinematic analysis.. Joints. 2016. DOI: 10.11138/jts/2016.4.2.121.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.