DOI 10.1515/bmt-2014-0181

Медицинская библиотека · Biomedizinische Technik. Biomedical engineering · 2016

Initial experience with imaging of the lower extremity arteries in an open 1.0 Tesla MRI system using the triggered angiography non-contrast-enhanced sequence (TRANCE) compared to digital subtraction angiography (DSA).

Suttmeyer Britta, Teichgräber Ulf, Rathke Hendrik, Albrecht Liane, Güttler Felix, Schnackenburg Bernhard, Hamm Bernd, de Bucourt Maximilian

СтатьяПеревод в очереди

Аннотация

PURPOSE: The aim of this study was to evaluate the feasibility and validity of arterial lower limb imaging with triggered angiography non-contrast enhanced (TRANCE) in an open MRI at 1.0 Tesla (T) compared to digital subtraction angiography (DSA).

MATERIAL AND

METHODS

ECG-gated, non-contrast-enhanced magnetic resonance angiography (MRA) was performed in a 1.0-T high-field open magnetic resonance imaging (MRI) system which generates a vertical magnetic field. Three acquisition levels were defined (abdominal and pelvic level, arterial segments above the knee and segments below the knee) and a total of 1782 vessel diameter measurements were taken on a total of 11 patients with suspected peripheral arterial occlusive disease (PAOD) (8 men, 3 women; average age 66 years). In each patient, 162 vessel segments (81 each with TRANCE and DSA) were defined and measured. Pearson correlation coefficients were calculated.

RESULTS

At the abdominal/pelvic level, all mean values measured with DSA exceeded the mean values obtained with TRANCE. Above the knee, mean vessel diameters were measured smaller in DSA in six, equal in three, and larger in two vessel segments. Below the knee, all measured averages, except for the tibiofibular tract (TFT) measurements, were larger in TRANCE. In total, two small (≤0.3), two moderate (>0.3), 11 good (>0.5), 10 high (>0.7) and 13 very high (>0.8) correlations were obtained.

CONCLUSIONS

Non-contrast-enhanced imaging of the lower limb arteries using a TRANCE-sequence in a 1.0 T open MRI system is feasible with the protocol presented; however, TRANCE tends to underestimate larger vessels and overestimate smaller vessels compared to DSA.

Для цитирования: Suttmeyer Britta, Teichgräber Ulf, Rathke Hendrik, Albrecht Liane, Güttler Felix, Schnackenburg Bernhard, Hamm Bernd, de Bucourt Maximilian. Initial experience with imaging of the lower extremity arteries in an open 1.0 Tesla MRI system using the triggered angiography non-contrast-enhanced sequence (TRANCE) compared to digital subtraction angiography (DSA).. Biomedizinische Technik. Biomedical engineering. 2016. DOI: 10.1515/bmt-2014-0181.