Revascularization for critical limb ischemia using the SpiderFX embolic protection device in the below-the-knee circulation: initial results.

СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Ward Thomas J, Piechowiak Rachel L, Patel Rahul S, Fischman Aaron M, Nowakowski F Scott, Kim Edward, Ellozy Sharif H, Faries Peter L, Lookstein Robert A

Аннотация

PURPOSE: To examine the safety and efficacy of the SpiderFX embolic protection device (EPD) in the below-the-knee (BTK) circulation in patients with critical limb ischemia (CLI).

MATERIALS AND METHODS: A single-center retrospective review was performed to identify patients with CLI and single-vessel runoff in whom the SpiderFX EPD was used in the BTK circulation. Technical success and device-related complications were the primary endpoints. Retrieval of macroscopic debris in the EPD, 1-year freedom from major adverse limb events (MALEs), and 30-day perioperative death were also evaluated. A major amputation, surgical bypass, endovascular thrombectomy, or endovascular thrombolysis was considered a MALE. Thirty-six patients (21 men; mean age, 75.8 y) treated between 2008 and 2013 had endovascular revascularization with use of the SpiderFX EPD in the BTK circulation.

RESULTS: The SpiderFX EPD was successfully deployed in all cases; the technical success rate of revascularization was 100%. Two minor and zero major complications were observed related to the SpiderFX. Two MALEs, a major amputation and a subsequent surgical bypass, were observed in the cohort. All MALEs occurred within 1 year of treatment (1-y freedom from MALE rate, 90%). Debris was retrieved in the SpiderFX device in 47% of patients.

CONCLUSIONS: The use of the SpiderFX EPD in the BTK circulation in patients with CLI is safe and frequently retrieves debris.