DOI 10.1002/jor.20778

Медицинская библиотека · Journal of orthopaedic research : official publication of the Orthopaedic Research Society · 2009

Noninvasive monitoring of elevated intramuscular pressure in a model compartment syndrome via quantitative fascial motion.

Lynch John E, Lynch John K, Cole Steven L, Carter Jonathan A, Hargens Alan R

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

Compartment syndromes, conditions of elevated intramuscular pressure (IMP) resulting from trauma or chronic overuse, frequently require invasive IMP monitoring for accurate diagnosis. Our objective was to test a noninvasive ultrasound technique for estimating IMP based on fascial displacement waveforms from arterial blood pressure pulses. IMP was increased in the legs of 23 healthy adult subjects up to 80 mmHg using two blood pressure cuffs covering the region from the knee to the ankle. Receiver operator characteristic curves and recursive partitioning were used to determine the sensitivity and specificity of diagnosing elevated IMP using fascial displacement. For one curve, in which several ultrasonic measurement parameters were used along with subject body mass index and blood pressure, the sensitivity and specificity for diagnosing normal IMP (below 30 mmHg) from elevated IMP (30 mmHg and up) was 0.61 and 0.94, respectively. Recursive partitioning, in which IMP was divided into three ranges (normal <30 mmHg, midrange of 30-40 mmHg, and elevated >or=50 mmHg), resulted in improved diagnostic sensitivity (0.77) with almost no change in specificity (0.93).

Для цитирования: Lynch John E, Lynch John K, Cole Steven L, Carter Jonathan A, Hargens Alan R. Noninvasive monitoring of elevated intramuscular pressure in a model compartment syndrome via quantitative fascial motion.. Journal of orthopaedic research : official publication of the Orthopaedic Research Society. 2009. DOI: 10.1002/jor.20778.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.