Contrast-enhanced MR mammography for evaluation of the contralateral breast in patients with diagnosed unilateral breast cancer or high-risk lesions.

Pediconi Federica, Catalano Carlo, Roselli Antonella, Padula Simona, Altomari Fiorella, Moriconi Enrica, Pronio Anna Maria, Kirchin Miles A, Passariello Roberto

Оценочное исследованиеСтатьяПеревод в очереди

Аннотация

PURPOSE: To prospectively evaluate accuracy of gadobenate dimeglumine-enhanced magnetic resonance (MR) mammography for depiction of synchronous contralateral breast cancer in patients with newly diagnosed unilateral breast cancer or high-risk lesions, with histologic analysis or follow-up as reference.

MATERIALS AND

METHODS

The study had ethics committee approval; all patients provided written informed consent. One hundred eighteen consecutive women (mean age, 52 years) with unilateral breast cancer or high-risk lesions and negative findings in the contralateral breast at physical examination, ultrasonography, and conventional mammography underwent gadobenate dimeglumine-enhanced 1.5-T MR mammography. Transverse three-dimensional T1-weighted gradient-echo images were acquired before and at 0, 2, 4, 6, and 8 minutes after gadobenate dimeglumine administration (0.1 mmol per kilogram body weight). Breast Imaging Reporting and Data System (BI-RADS) was used to categorize breast density and the level of suspicion for malignant contralateral breast lesions. Results were compared with histologic findings. Sensitivity, specificity, accuracy, and positive and negative predictive values for contrast-enhanced MR mammography were evaluated.

RESULTS

Contrast-enhanced MR mammography revealed contralateral lesions in 28 (24%) of 118 patients. Twenty-four lesions were detected in patients with dense breasts (BI-RADS breast density category III or IV). Lesions in eight (29%) of 28 patients were BI-RADS category 4; patients underwent biopsy. Lesions in 20 (71%) patients were BI-RADS category 5; patients underwent surgery. At histologic analysis, 22 lesions were confirmed as malignant; six lesions were fibroadenomas. No false-negative lesions were detected; none of the fibroadenomas were BI-RADS category 5. The sensitivity, specificity, accuracy, and positive and negative predictive values of contrast-enhanced MR mammography for depiction of malignant or high-risk contralateral lesions were 100%, 94%, 95%, 79%, and 100%, respectively. Follow-up findings (12-24 months) confirmed absence of contralateral lesions in 90 of 118 patients with negative contrast-enhanced MR mammographic findings in the contralateral breast.

CONCLUSION

Contrast-enhanced MR mammography is accurate for detection of synchronous contralateral cancer or high-risk lesions in patients with newly diagnosed breast cancer or high-risk lesions.

Для цитирования: Pediconi Federica, Catalano Carlo, Roselli Antonella, Padula Simona, Altomari Fiorella, Moriconi Enrica, Pronio Anna Maria, Kirchin Miles A, Passariello Roberto. Contrast-enhanced MR mammography for evaluation of the contralateral breast in patients with diagnosed unilateral breast cancer or high-risk lesions.. Radiology. 2007. DOI: 10.1148/radiol.2433060838.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.