Analytical agreement and clinical interchangeability of routine complete blood count parameters between the Atellica HEMA 580 and Sysmex XN-1000 analyzers: a CLSI EP09c study.

Nguyen Quang Tung, Pham Van Tuan, Nguyen Chi Thanh, Nguyen Phuc Minh Hoang, Nguyen Dieu Huong, Tong Huong Lan

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Analytical consistency in complete blood count (CBC) testing is essential when introducing new hematology analyzers into routine clinical use. This study evaluated the agreement between the Atellica HEMA 580 and the Sysmex XN-1000 using a standardized method-comparison approach.

METHODS

A total of 100 residual K₂EDTA whole-blood samples were analyzed on both systems between April and June 2025. Method comparison followed CLSI EP09c guidelines, including Passing-Bablok regression, Spearman correlation, bias estimation at predefined medical-decision levels, and Bland-Altman analysis.

RESULTS

Strong analytical agreement was observed for most parameters, with correlation coefficients exceeding 0.98 for key indices including RBC (r = 0.999), HGB (r = 0.998), WBC (r = 0.996), and PLT-I (r = 0.995), and regression slopes close to unity. Bias estimates at medical-decision levels were within allowable limits for the majority of parameters. WBC at 0.5 × 109/L and PLT-I at 100 × 109/L achieved Outcome A, indicating optimal agreement at clinically critical thresholds. In contrast, MCV showed proportional deviation with bias exceeding allowable limits at both 80 fL and 100 fL, and RDW-CV demonstrated lower correlation (r = 0.848) and greater dispersion. Bland-Altman analysis confirmed small mean bias for most parameters, with narrow limits of agreement for RBC (-1.48%, LoA - 4.19 to 1.36) and HGB (1.54%, LoA - 1.37 to 4.19), while wider limits were observed for WBC (-12.62 to 1.70) and PLT-I (-16.55 to 9.51). RDW-CV exhibited the broadest limits of agreement (-17.76 to 10.38). Among differential parameters, lymphocyte count showed optimal agreement (bias -1.12%, Outcome A), while other indices remained within acceptable ranges.

CONCLUSION

The Atellica HEMA 580 shows acceptable analytical agreement with the Sysmex XN-1000 for RBC, HGB, WBC, PLT-I, and the evaluated neutrophil and lymphocyte differential parameters, and can be applied in clinical laboratory practice. In contrast, MCV and RDW-CV demonstrated analyzer-dependent differences that warrant cautious interpretation when comparing results across platforms or during longitudinal assessment.

Для цитирования: Nguyen Quang Tung, Pham Van Tuan, Nguyen Chi Thanh, Nguyen Phuc Minh Hoang, Nguyen Dieu Huong, Tong Huong Lan. Analytical agreement and clinical interchangeability of routine complete blood count parameters between the Atellica HEMA 580 and Sysmex XN-1000 analyzers: a CLSI EP09c study.. Frontiers in medicine. 2026. DOI: 10.3389/fmed.2026.1878534.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.