Real-world Glycemic Outcomes of a Tubeless Automated Insulin Delivery System (Medtrum A8 TouchCare) in Adults With Type 1 Diabetes.

Modestino Michele Roberto, De Luca Mariarosaria, De Fortuna Umberto, Ferrentino Laura, Iacono Olimpia, Verdoliva Rita, Foglia Angelo, Guardasole Vincenzo

СтатьяНаблюдательное исследованиеПеревод пока не подготовлен

Аннотация

OBJECTIVES

Automated insulin delivery (AID) systems are recommended as a first-line therapy for type 1 diabetes. The Medtrum A8 TouchCare is a tubeless AID system powered by the APGO algorithm, which modulates basal insulin and delivers automatic correction microboluses. Real-world evidence on this system remains extremely limited. In this study, we aimed to evaluate the efficacy of the Medtrum A8 TouchCare in AID mode in a real-world cohort of adults with type 1 diabetes.

METHODS

We conducted a single-centre, observational retrospective study at Federico II University Hospital, Naples, Italy (January to October 2025). Adults with type 1 diabetes of ≥6 months' duration who were using the Medtrum A8 TouchCare in AID mode were included. Continuous glucose monitoring metrics and clinical parameters were compared at baseline and after 3 months. The primary endpoint was the change in time in range.

RESULTS

A total of 20 patients were included (mean age: 39±6 years; diabetes duration: 18±6 years). After 3 months, time in range increased by 12.51% (p<0.05); time above range 1 and time above range 2 decreased by 7.2% (p<0.05) and 5.97% (p<0.05), respectively. Glucose monitoring index improved by 0.46% (p<0.05) and glycemia risk index by 13.75 points (p<0.05), and glycated hemoglobin decreased by 0.52% (p<0.05). No significant changes were observed in time below range 1, time below range 2, or coefficient variability, and no severe hypoglycemia episodes occurred.

CONCLUSIONS

The Medtrum A8 TouchCare in AID mode significantly improved glycemic management without increasing hypoglycemia. These findings support the efficacy of this tubeless AID system and warrant confirmation in larger prospective studies.

Для цитирования: Modestino Michele Roberto, De Luca Mariarosaria, De Fortuna Umberto, Ferrentino Laura, Iacono Olimpia, Verdoliva Rita, Foglia Angelo, Guardasole Vincenzo. Real-world Glycemic Outcomes of a Tubeless Automated Insulin Delivery System (Medtrum A8 TouchCare) in Adults With Type 1 Diabetes.. Canadian journal of diabetes. 2026. DOI: 10.1016/j.jcjd.2026.05.004.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.