MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1186/s41182-025-00891-6

Медицинская библиотека · Tropical medicine and health · 2026

Development of a colorimetric RT-LAMP-based microfluidic device for SARS-CoV-2 detection.

Sakurai Tatsuya, Aonuma Hiroka, Natsuhara Daigo, Hoshina Tokio, Ote Manabu, Saiki Erisha, Shibata Takayuki, Kanuka Hirotaka

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Diagnosing COVID-19, caused by the coronavirus SARS-CoV-2, mainly involves detecting specific viral gene sequences. To effectively control highly contagious diseases, such as COVID-19, it is essential to develop quick, accurate, sensitive, and easy-to-use diagnostic methods for medical settings. The current gold standard, reverse transcription-quantitative polymerase chain reaction (RT-qPCR), requires specialized equipment and trained personnel and is typically conducted in dedicated laboratories. Loop-mediated isothermal amplification (LAMP) provides a rapid and simple alternative under isothermal conditions. In this study, we developed a colorimetric reverse transcription-loop-mediated isothermal amplification (RT-LAMP) assay for SARS-CoV-2 detection and integrated it into a microfluidic device.

METHODS

RT-LAMP primers targeting conserved regions of the SARS-CoV-2 envelope (E) and RNA-dependent RNA polymerase (RdRp) genes were designed. Hydroxy naphthol blue (HNB) was used as a color indicator for visual detection. The microfluidic device, fabricated from polydimethylsiloxane (PDMS), contained five reaction chambers. RNA samples extracted from nasopharyngeal swabs of COVID-19 patients were tested using the RT-LAMP microfluidic device. Detection performance of the device was assessed using RT-qPCR as the reference standard.

RESULTS

Among 86 RNA samples, 66 were RT-qPCR positive and 20 were negative. The RT-LAMP microfluidic device achieved 100% specificity (95% CI 83.2-100%) and an overall sensitivity of 75.8% (95% CI 63.6-85.5%) for samples with Ct ≤ 35. Sensitivity correlated with viral load, showing 100% for samples with Ct ≤ 25, 88.6% for Ct ≤ 30, and 50% for 30 < Ct ≤ 35.

CONCLUSIONS

The colorimetric RT-LAMP-based microfluidic device enables simple, rapid, and reliable detection of SARS-CoV-2 RNA with high specificity and sensitivity. The device enables visual detection of results and requires only simple equipment, making it suitable as a practical molecular detection tool for point-of-care and resource-limited settings. In terms of detection performance, our findings, when considered alongside previously published data, indicate that the sensitivity and specificity of the device are comparable to or exceed those of lateral flow tests commonly used for point-of-care COVID-19 diagnostics.

Для цитирования: Sakurai Tatsuya, Aonuma Hiroka, Natsuhara Daigo, Hoshina Tokio, Ote Manabu, Saiki Erisha, Shibata Takayuki, Kanuka Hirotaka. Development of a colorimetric RT-LAMP-based microfluidic device for SARS-CoV-2 detection.. Tropical medicine and health. 2026. DOI: 10.1186/s41182-025-00891-6.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.