DOI 10.1007/s00216-023-04518-9
Development of highly accurate digital PCR method and reference material for monkeypox virus detection.
Аннотация
Human monkeypox has attracted attention recently. Monkeypox virus (MPXV) keeps evolving as it spreading around the world rapidly, which may threaten the health of more and more people. Here, we have developed a high order reference method based on digital PCR (dPCR) for MPXV detection, of which the limits of quantification (LoQ) and detection (LoD) are 38 and 6 copies/reaction, respectively. Pseudovirus reference materials (RM) containing the conserved F3L gene has been developed, and the homogeneity assessment showed that the RM was homogeneous. The reference value with its expanded uncertainty determined by the established dPCR is (2.74 ± 0.46) × 103 copies/μL. Six different MPXV test kits were accessed by the RM. Four out of six test kits cannot reach their claimed LoDs. The poor analytical sensitivity might cause false-negative results, which lead to incorrect diagnosis and treatment. The establishment of a high order reference method of dPCR and pseudovirus RM is very useful for improving the accuracy and reliability of MPXV detection.