OBJECTIVE

To compare the temperature changes with laser activation when providing irrigation with a pressure bag vs a fluid management system (FMS).

MATERIALS AND

METHODS

Benchtop experiments were performed with a LithoVue™ Elite (Boston Scientific Corp., Marlborough, MA, USA) flexible ureteroscope placed through an 11/13-F ureteric access sheath into a benchtop pelvicalyceal kidney model. Room temperature saline irrigation, via the scope, was provided either with a pressure bag or FMS. Both systems were pressurised to 200 mmHg and left for 10 min. After 10 min, a holmium:yttrium-aluminium-garnet laser was activated at 20, 40, or 60 W for 90 s. Flow rate, irrigation pressure, intrarenal pressure and calyceal temperature were recorded continuously. Experiments were repeated five times. Benchtop empirical temperature data were compared to mathematical model predictions. The mathematical model was then used to predict calyx temperature as a function of flow rate in in vivo conditions.

RESULTS

Pressure-bag irrigation flow rate declined from ~38 to ~25 mL/min (36%) over 10 min, whereas the FMS maintained stable flow of ~45 mL/min. Intrarenal pressure remained ≤30 mmHg in all experiments. Consequently, the benchtop calyx mean (standard deviation) temperature rise with pressure bag vs FMS irrigation was: at 20 W 7.9 (1.0)°C vs 6.4 (0.6)°C (P = 0.02); at 40 W 12.8 (3.2)°C vs 10.3 (1.7)°C (P = 0.15); at 60 W 17 (3.4)°C vs 12.4 (1.2)°C (P = 0.02).

CONCLUSION

Irrigation flow rate is a key determinant of thermal risk during laser lithotripsy. An automated FMS can maintain consistent flow; mitigating temperature rise and thermal dose in a benchtop model.

Для цитирования: Menzies-Wilson Richard, Williams Jessica, Rhodes Candace, Glienke Leilane, Roberts William W, Turney Ben. Automated irrigation systems in mitigating laser-induced temperature during simulated ureteroscopy.. BJU international. 2026. DOI: 10.1111/bju.70364.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.