MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1093/eschf/xvag070

Медицинская библиотека · ESC heart failure · 2026

The ALT FLOW early feasibility study 3-year results assessing left atrial to coronary sinus shunting in HFpEF.

Fioretti Francesco, Hibbert Benjamin, Eckman Peter M, Simard Trevor, Labinaz Marino, Nazer Babak, Wiley Mark, Gupta Bhanu, Sauer Andrew J, Shah Hirak, Sorajja Paul, Pineda Andres M, Missov Emil, Aldaia Lillian, Koulogiannis Konstantinos, Tedford Ryan J, Gray William A, Zahr Firas, Butler Javed

Клиническое исследованиеСтатьяМногоцентровое исследованиеПеревод пока не подготовлен

Аннотация

INTRODUCTION

Left atrial-to-coronary sinus (LA-to-CS) shunting through the APTURE transcatheter shunt system aims to unload the LA while preserving the interatrial septum. This article aims to examine the 3-year health status, functional capacity, and clinical outcomes of the ALT-FLOW Early Feasibility Study (EFS).

METHODS

Device safety, patency, and echocardiographic parameters, but also symptoms, health status, and exercise capacity were assessed in patients with left ventricular ejection fraction (LVEF) > 40% (n & 95). Meta-Analysis Global Group in Chronic Heart Failure (MAGGIC) score predicted 3-year all-cause mortality was compared with observed mortality assessed by Kaplan-Meier analysis using adjudicated events. Predicted and observed HF hospitalization (HFH)-free survival rates were also compared.

RESULTS

At 3 years, the safety profile remained stable with no new safety signals, and all implanted shunts remained patent. 76.3% of patients improved by at least one NYHA class, and mean KCCQ Overall Summary Scores increased by 25 points from baseline (P < .0001). The 6-min walk distance improved by an average of 39 m (P & .002). These benefits were consistent regardless of baseline pulmonary vascular resistance (PVR). Right heart echocardiographic parameters after 3 years were stable, though a reduction in TAPSE was observed, primarily in the high PVR subgroup. The observed 3-year survival was significantly higher than the MAGGIC-predicted survival (84.4% vs 70.2, respectively; P & .009), as well as the HFH-free survival (68.0% vs 24.0%, respectively; P < 0001).

CONCLUSION

In patients with HF and LVEF >40%, the APTURE LA-to-CS shunting confirmed an acceptable safety profile with stable right heart parameters and sustained improvements in symptoms, quality of life, and functional capacity at 3 years. Observed survival and HFH were superior to MAGGIC predictions.

Для цитирования: Fioretti Francesco, Hibbert Benjamin, Eckman Peter M, Simard Trevor, Labinaz Marino, Nazer Babak, Wiley Mark, Gupta Bhanu, Sauer Andrew J, Shah Hirak, Sorajja Paul, Pineda Andres M, Missov Emil, Aldaia Lillian, Koulogiannis Konstantinos, Tedford Ryan J, Gray William A, Zahr Firas, Butler Javed. The ALT FLOW early feasibility study 3-year results assessing left atrial to coronary sinus shunting in HFpEF.. ESC heart failure. 2026. DOI: 10.1093/eschf/xvag070.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.