Disposable pressure transducer to identify central pressure measurements in umbilical lines for preterm and term infants in the neonatal intensive care unit.

Moore Carrie, Devlin Lori A, Crabtree Cynthia, Farris Amanda, Kurtz Joshua, Marques Keri

СтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVES

Assess the feasibility of using a disposable pressure transducer with integrated digital display to differentiate arterial versus venous line placement in neonates.

STUDY

DESIGN

Infants ≥23 weeks' gestation with appropriately placed umbilical catheters were enrolled. A single baseline pressure reading was obtained, hypothesizing that arterial placement could be differentiated with a pressure ≥12 mmHg.

RESULTS

Forty-five infants were enrolled to obtain 62 measurements, 31 from both umbilical artery catheters(UAC) and umbilical venous catheters(UVC). 100% of UAC were ≥12 mmHg and 87% of UVC were <12 mmHg. Median device pressures for UAC and UVC were 41 mmHg (IQR 31-45 mmHg) and 5 mmHg (IQR 3-9 mmHg) (p-value < 0.0001). An optimal venous device pressure cut point was determined to be 21 mmHG (97% sensitivity, 100% specificity, AUC 0.98).

CONCLUSION

This transducer is likely safe and can differentiate line placement in arterial and venous systems. 12 mmHg appears to reliably identify lines placed in the arterial system, but 21 mmHG was found to be the optimal venous cut point.

Для цитирования: Moore Carrie, Devlin Lori A, Crabtree Cynthia, Farris Amanda, Kurtz Joshua, Marques Keri. Disposable pressure transducer to identify central pressure measurements in umbilical lines for preterm and term infants in the neonatal intensive care unit.. Journal of perinatology : official journal of the California Perinatal Association. 2025. DOI: 10.1038/s41372-025-02338-6.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.