Derivo embolization device for intracranial aneurysms: a Spanish multicenter retrospective study.

Piñana Carlos, Remollo Sebastian, Zamarro Joaquín, Werner Mariano, Espinosa de Rueda Mariano, Vega Pedro, Hernandez David, Murias Eduardo, Rivera Eila, Olier Jorge, San Roman Luis, Páez-Carpio Alfredo, Requena Manuel, Aixut Sonia, de Dios Lascuevas Marta, Moreu Manuel, Rosati Santiago, Gramegna Laura Ludovica, Castaño Carlos, Tomasello Alejandro

Многоцентровое исследованиеСтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Flow diverters have emerged in recent years as a safe and effective treatment for intracranial aneurysms, with expanding indications. The Derivo embolization device (DED) is a second-generation flow diverter with a surface finish that may reduce thrombogenicity. We report our multicenter experience evaluating its safety and efficacy.

METHODS

We retrospectively analyzed all patients treated with the DED in eight centers in Spain between 2016 and 2020. Demographics, clinical data, procedural complications, morbidity and aneurysm occlusion rates were collected.

RESULTS

A total of 209 patients with 250 aneurysms were treated (77.5% women). The majority of aneurysms were located in the internal carotid artery (86.8%) and most (69.2%) were small (<10 mm) with a median maximum diameter of 5.85 mm and median neck size of 4 mm. DED deployment was successful in all cases, despite two malfunctioning devices (1%). Major complications occurred in nine patients (4.3%), while mild neurologic clinical events were registered in 23 (11%); four patients died (1.9%). A total of 194 aneurysms had an angiographic follow-up at 6 months and showed complete aneurysm occlusion in 75% of cases. Twelve-month follow-up was available for 112 of the treated aneurysms, with a total occlusion rate of 83%.

CONCLUSION

The DED is a second-generation surface-modified flow diverter that presents an option for treatment of intracranial aneurysms with comparable safety and efficacy to other available flow diverter devices. Nonetheless, risks are not negligible, and must be balanced against the natural history risk of cerebral aneurysms, considering the tendency to widen indications for treatment of smaller and less complex lesions in day-to-day use.

Для цитирования: Piñana Carlos, Remollo Sebastian, Zamarro Joaquín, Werner Mariano, Espinosa de Rueda Mariano, Vega Pedro, Hernandez David, Murias Eduardo, Rivera Eila, Olier Jorge, San Roman Luis, Páez-Carpio Alfredo, Requena Manuel, Aixut Sonia, de Dios Lascuevas Marta, Moreu Manuel, Rosati Santiago, Gramegna Laura Ludovica, Castaño Carlos, Tomasello Alejandro. Derivo embolization device for intracranial aneurysms: a Spanish multicenter retrospective study.. Journal of neurointerventional surgery. 2022. DOI: 10.1136/jnis-2022-019220.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.