Novel Fluoroscopic Evaluation of Incorrect Impella Placement.

Hill Andrew P, Slack Michael, Waksman Ron, Ben-Dor Itsik

СтатьяКлинический случайVideo-Audio MediaПеревод пока не подготовлен

Аннотация

A 65-year-old man with newly diagnosed cardiomyopathy was found to have multivessel coronary artery disease. After he was deemed not to be a surgical candidate, Impella-supported percutaneous coronary intervention was successfully performed on his left main and left circumflex arteries. Although initially thought to be in the correct position by standard single plane fluoroscopy, the Impella catheter was shown with the use of additional non-standard fluoroscopic views to be oriented toward the mitral apparatus. Malrotation of the Impella microaxial flow pump catheter can lead to complications and suboptimal unloading. This case highlights a novel fluoroscopic technique for evaluation of incorrect Impella placement.

Для цитирования: Hill Andrew P, Slack Michael, Waksman Ron, Ben-Dor Itsik. Novel Fluoroscopic Evaluation of Incorrect Impella Placement.. Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions. 2025. DOI: 10.1002/ccd.31651.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.