MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1093/ehjci/jeae236

Медицинская библиотека · European heart journal. Cardiovascular Imaging · 2024

Reproducibility of cardiac computed tomography classifications of hypoattenuated thickening and peridevice leak following left atrial appendage closure.

Kramer Anders, Lo Russo Gerardo, Alarouri Hasan S, Collins Jeremy D, Møller Jensen Jesper, Nielsen-Kudsk Jens Erik, Alkhouli Mohamad, Korsholm Kasper

СтатьяМногоцентровое исследованиеПеревод пока не подготовлен

Аннотация

AIMS: To assess the reproducibility of interpreting hypoattenuated thickening (HAT) and peridevice leak (PDL) using cardiac computed tomography (CT) imaging following Watchman FLX left atrial appendage closure (LAAC).

METHODS AND

RESULTS

In this multicentre retrospective reproducibility study, 100 anonymized post-LAAC cardiac CT scans were evaluated within the same cardiac phase by an experienced and a novice rater blinded to prior evaluations. All scans were evaluated twice by each rater, assessing overall HAT and PDL categories as well as specific associated findings based on suggested algorithms for post-LAAC interpretation. Inter- and intra-rater agreement and reliability were evaluated using absolute agreement, Cohen's kappa, and Kendall's tau for categorical variables, and mean difference, Bland-Altman plots, limits of agreement (lower and upper), and intra-class correlation coefficients (ICCs) for continuous variables. Within overall categories of both HAT and PDL, substantial agreement (kappa ≥0.61) and reliability (Kendall's tau-b ≥ 0.75) were observed. Specifically, identifying high-grade HAT (kappa ≥0.78) and distal patency (kappa ≥0.85) displayed the highest agreement within HAT and PDL interpretation. Meanwhile, measuring the height of the proximal screw hub cove represented the least reliable HAT assessment among both inter- and intra-rater comparisons (ICC < 0.75), while suspected leak mechanism represented the least reproducible PDL measure.

CONCLUSION

Despite only minimal training of one rater, overall high levels of inter- and intra-rater agreement and reliability were observed across the chosen algorithms for interpretation of HAT and PDL following Watchman FLX LAAC. The prognostic implications of the included variables are to be explored in future trials and registries.

Для цитирования: Kramer Anders, Lo Russo Gerardo, Alarouri Hasan S, Collins Jeremy D, Møller Jensen Jesper, Nielsen-Kudsk Jens Erik, Alkhouli Mohamad, Korsholm Kasper. Reproducibility of cardiac computed tomography classifications of hypoattenuated thickening and peridevice leak following left atrial appendage closure.. European heart journal. Cardiovascular Imaging. 2024. DOI: 10.1093/ehjci/jeae236.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.