DOI 10.1038/s41598-024-62440-1
Effect of prolonged pressure equalization on final drifting during pressure wire studies.
Аннотация
Pressure drifting is a troublesome error in invasive coronary function tests. This study aimed to evaluate the relationship between prolonged and short-time pressure equalizations in pressure drifting. Pressure drifting was defined as the pressure gradient between the mean pressure of the distal wire sensor (Pd) and aortic pressure (Pa) when the wire was withdrawn to the tip of the guiding catheter. Significant drifts 1 and 2 were defined as the absolute values of pressure gradients > 2 and > 3 mmHg, respectively. A logistic regression model was used to evaluate the associations between prolonged pressure equalization and each pressure drifting. The prolonged pressure equalization strategy was associated with a lower incidence of drift 1 than the short-time pressure equalization strategy (6.84% vs. 16.92%, p < 0.05). However, no statistical differences were found in the incidence of drift 2 between the prolonged and short-time pressure equalization strategies (4.27% vs. 7.69%, p = 0.34). In the multivariable regression model, only the prolonged pressure equalization strategy predicted a lower incidence of pressure drift 1. In conclusion, the prolonged pressure equalization strategy was associated with a lower incidence of significant pressure drifting with more stringent thresholds than the short-time pressure equalization strategy.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.