Nakashima Masaki, Yamamoto Masanori, Sago Mitsuru, Tanaka Shuhei, Chatani Ryuki, Asami Masahiko, Hachinohe Daisuke, Naganuma Toru, Ohno Yohei, Tani Tomoyuki, Okamatsu Hideharu, Mizutani Kazuki, Watanabe Yusuke, Izumo Masaki, Saji Mike, Mizuno Shingo, Ueno Hiroshi, Kubo Shunsuke, Shirai Shinichi, Hayashida Kentaro, OCEAN-LAAC Investigators

СтатьяМногоцентровое исследованиеСравнительное исследованиеПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

BACKGROUND

Limited data are available regarding clinical outcomes after percutaneous left atrial appendage closure using WATCHMAN FLX (WM-FLX) and WATCHMAN-2.5 (WM2.5) devices in Asian patients.

METHODS AND

RESULTS

Data of 1,464 consecutive patients (WM-FLX, n=909; WM2.5, n=555) were extracted from a Japanese multicenter registry, and clinical data were compared between the 2 groups. No in-hospital deaths, periprocedural stroke, or device embolization occurred. Procedural success was significantly higher in the WM-FLX than WM2.5 group (95.8% vs. 91.9%; P=0.002) owing to the lower incidence of periprocedural pericardial effusion (0.55% vs. 1.8%; P=0.021). No significant differences in all-cause death, postprocedural stroke, and device-related thrombus were observed between the 2 groups. However, the cumulative bleeding rate at 1 year was substantially lower in the WM-FLX group (7.8% vs. 16.4%; P<0.001). Landmark analysis of bleeding events highlighted lower bleeding rates in the WM-FLX than WM2.5 group within the first 6 months (6.4% vs. 14.8%; P<0.001), with comparable bleeding rates over the 6- to 12-month period (1.5% vs. 3.2%, respectively; P=0.065).

CONCLUSIONS

This study demonstrated higher early safety and lower 1-year bleeding rates in the WM-FLX than WM2.5 group. The lower bleeding events with WM-FLX are likely due to multiple factors other than purely difference in devices, such as postprocedural drug regimen.

Для цитирования: Nakashima Masaki, Yamamoto Masanori, Sago Mitsuru, Tanaka Shuhei, Chatani Ryuki, Asami Masahiko, Hachinohe Daisuke, Naganuma Toru, Ohno Yohei, Tani Tomoyuki, Okamatsu Hideharu, Mizutani Kazuki, Watanabe Yusuke, Izumo Masaki, Saji Mike, Mizuno Shingo, Ueno Hiroshi, Kubo Shunsuke, Shirai Shinichi, Hayashida Kentaro, OCEAN-LAAC Investigators. Comparative Data of Procedural and Midterm Outcomes in Patients Who Underwent Percutaneous Left Atrial Appendage Closure Between the WATCHMAN FLX and WATCHMAN 2.5 Devices - Insight From the OCEAN-LAAC Registry.. Circulation journal : official journal of the Japanese Circulation Society. 2024. DOI: 10.1253/circj.cj-24-0062.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.