DOI 10.1371/journal.pone.0302793
Comparison of cardiac output estimates obtained from the Antares oscillometric pulse wave analysis algorithm and from Doppler transthoracic echocardiography.
Аннотация
BACKGROUND
In cardiology, cardiac output (CO) is an important parameter for assessing cardiac function. While invasive thermodilution procedures are the gold standard for CO assessment, transthoracic Doppler echocardiography (TTE) has become the established method for routine CO assessment in daily clinical practice. However, a demand persists for non-invasive approaches, including oscillometric pulse wave analysis (PWA), to enhance the accuracy of CO estimation, reduce complications associated with invasive procedures, and facilitate its application in non-intensive care settings. Here, we aimed to compare the TTE and oscillometric PWA algorithm Antares for a non-invasive estimation of CO.
METHODS
Non-invasive CO data obtained by two-dimensional TTE were compared with those from an oscillometric blood pressure device (custo med GmbH, Ottobrunn, Germany) using the integrated algorithm Antares (Redwave Medical GmbH, Jena, Germany). In total, 59 patients undergoing elective cardiac catheterization for clinical reasons (71±10 years old, 76% males) were included. Agreement between both CO measures were assessed by Bland-Altman analysis, Student's t-test, and Pearson correlations.
RESULTS
The mean difference in CO was 0.04 ± 1.03 l/min (95% confidence interval for the mean difference: -0.23 to 0.30 l/min) for the overall group, with lower and upper limits of agreement at -1.98 and 2.05 l/min, respectively. There was no statistically significant difference in means between both CO measures (P = 0.785). Statistically significant correlations between TTE and Antares CO were observed in the entire cohort (r = 0.705, P<0.001) as well as in female (r = 0.802, P<0.001) and male patients (r = 0.669, P<0.001).
CONCLUSIONS
The oscillometric PWA algorithm Antares and established TTE for a non-invasive estimation of CO are highly correlated in male and female patients, with no statistically significant difference between both approaches. Future validation studies of the Antares CO are necessary before a clinical application can be considered.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.