Comprehensive radiation shield minimizes operator radiation exposure in coronary and structural heart procedures.
Аннотация
OBJECTIVES
This study evaluated the efficacy of a novel comprehensive shield designed to minimize radiation exposure (RE) to Physicians performing coronary and structural heart procedures.
BACKGROUND
The Protego™ radiation shielding system (Image Diagnostics Inc., Fitchburg, Ma) is designed to provide comprehensive protection from RE and has been State certified sufficient to allow operators to perform procedures without orthopedically burdensome lead aprons.
METHODS
This single center two-group cohort study assessed the efficacy of this shield in a large number of cardiac procedures (coronary and structural), comparing operator RE compared to standard protection methods (personal lead apparel and "drop down" shield).
RESULTS
The Protego™ system reduced operator RE by 99 % compared to Standard Protection. RE was significantly lower at both "Head" level by thyroid median dose 0.0 (0.0, 0,0) vs 5.7 (2.9, 8.2) μSv (p < 0.001), as well as waist dose 0.0 (0.0, 0.0) vs 10.0 (5.0, 16.6) μSv (p < 0.001). "Zero" Total RE was documented by Raysafe™ in 64 % (n = 32) of TAVR cases and 73.2 % (n = 183) of the coronary cases utilizing Protego™. In contrast, standard protection did not achieve "Zero" exposure in a single case. These dramatic differences in RE were achieved despite higher fluoroscopy times in the Protego™ arm (11.9 ± 8.6 vs 14.3 ± 12.5 min, p = 0.015). Per case procedural exposure measured by Dose Area Product was higher in the Protego™ group compared to standard protection (115.4 ± 139.2 vs 74.9 ± 69.3, p < 0.001).
CONCLUSION
The Protego™ shield provides total body RE protection for operators performing both coronary and structural heart procedures. This shield allows procedural performance without the need for personal lead aprons and has potential to reduce catheterization laboratory occupational health hazards.