MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.hrthm.2017.05.038

Медицинская библиотека · Heart rhythm · 2017

Efficacy and safety of left atrial appendage closure with WATCHMAN in patients with or without contraindication to oral anticoagulation: 1-Year follow-up outcome data of the EWOLUTION trial.

Boersma Lucas V, Ince Hueseyin, Kische Stephan, Pokushalov Evgeny, Schmitz Thomas, Schmidt Boris, Gori Tommaso, Meincke Felix, Protopopov Alexey Vladimir, Betts Timothy, Foley David, Sievert Horst, Mazzone Patrizio, De Potter Tom, Vireca Elisa, Stein Kenneth, Bergmann Martin W, EWOLUTION Investigators

Клиническое исследованиеСтатьяМногоцентровое исследованиеПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

BACKGROUND

Left atrial appendage (LAA) occlusion with WATCHMAN has emerged as viable alternative to vitamin K antagonists in randomized controlled trials.

OBJECTIVE

EWOLUTION was designed to provide data in routine practice from a prospective multicenter registry.

METHODS

A total of 1025 patients scheduled for a WATCHMAN implant were prospectively and sequentially enrolled at 47 centers. Indication for LAA closure was based on European Society of Cardiology guidelines. Follow-up and transesophageal echocardiography (TEE) were performed per local practice.

RESULTS

The baseline CHA2DS2-VASc score was 4.5 ± 1.6; the mean age was 73.4 ± 9 years; previous transient ischemic attack/ischemic stroke was present in 312 (30.5%), 155 (15.1%) had previous hemorrhagic stroke, and 320 (31.3%) had a history of major bleeding; and 750 (73%) were deemed unsuitable for oral anticoagulation therapy. WATCHMAN implant succeeded in 1005 (98.5%) of patients, without leaks >5 mm in 1002 (99.7%) with at least 1 TEE follow-up in 875 patients (87%). Antiplatelet therapy was used in 784 (83%), while vitamin K antagonists were used in only 75 (8%). At 1 year, mortality was 98 (9.8%), reflecting the advanced age and comorbidities in this population. Device thrombus was observed in 28 patients at routine TEE (3.7%) and was not correlated with the drug regimen (P = .14). Ischemic stroke rate was 1.1% (relative risk 84% vs estimated historical data); the major bleeding rate was 2.6% and was predominantly (2.3%) nonprocedure/device related.

CONCLUSION

LAA closure with the WATCHMAN device has a high implant and sealing success. This method of stroke risk reduction appears to be safe and effective with an ischemic stroke rate as low as 1.1%, even though 73% of patients had a contraindication to and were not using oral anticoagulation.

Для цитирования: Boersma Lucas V, Ince Hueseyin, Kische Stephan, Pokushalov Evgeny, Schmitz Thomas, Schmidt Boris, Gori Tommaso, Meincke Felix, Protopopov Alexey Vladimir, Betts Timothy, Foley David, Sievert Horst, Mazzone Patrizio, De Potter Tom, Vireca Elisa, Stein Kenneth, Bergmann Martin W, EWOLUTION Investigators. Efficacy and safety of left atrial appendage closure with WATCHMAN in patients with or without contraindication to oral anticoagulation: 1-Year follow-up outcome data of the EWOLUTION trial.. Heart rhythm. 2017. DOI: 10.1016/j.hrthm.2017.05.038.