Phantom Study of Arterial Localization using Tactile Sensor Array and a Normal Vs. Shear Pulse Pressure Propagation Method.

Hampson Rory, Lawley Alistair, Dobie Gordon

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

OBJECTIVE

Locating the radial artery reliably is a key challenge in reducing patient risks from complications in Trans-Radial Access, which is an important clinical method for catheterization, cardiac monitoring, and neuroendovascular procedures. New tactile sensing technology is being developed to bridge the skill, cost, and performance gap between ultrasonic needle guidance, and manual palpation, for use in developing countries. This paper further develops tactile artery localization with a novel algorithm for arterial localization based on the properties of a curved tactile sensor array.

METHODS

Using tactile sensor insensitivity to shear loading, coupled with a radial pulse wave propagation path, the position of the artery can be found at the intersection of a normal and tangential vector from the array corresponding to maximum and minimum pulse pressure measurement locations respectively. This was validated in a simple silicone phantom study

Results

The proposed method measured with MAE= 0.58±0.25mm whilst the artery is within range of the tactile array, compared with 0.81±0.57mm for a comparative method of simple pulse localization. This showed improvement in arterial localization and repeatability, and was within 1 arterial radius, expected to reduce the risk of missing the artery, or perforating the side wall.Clinical Relevance- Robust and repeatable arterial localization is important for reducing the failure rate of trans-radial (and other arterial) procedures, and thus reducing the risk of harmful complications.

Для цитирования: Hampson Rory, Lawley Alistair, Dobie Gordon. Phantom Study of Arterial Localization using Tactile Sensor Array and a Normal Vs. Shear Pulse Pressure Propagation Method.. Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual International Conference. 2023. DOI: 10.1109/embc40787.2023.10340678.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.