DOI 10.1093/europace/euad316

Медицинская библиотека · Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology · 2023

Comparison of non-laser and laser transvenous lead extraction: a systematic review and meta-analysis.

Akhtar Zaki, Kontogiannis Christos, Georgiopoulos Georgios, Starck Christoph T, Leung Lisa W M, Lee Sun Y, Lee Byron K, Seshasai Sreenivasa R K, Sohal Manav, Gallagher Mark M

МетаанализСистематический обзорСтатьяПеревод пока не подготовлен

Аннотация

AIMS: Transvenous lead extraction (TLE) is performed using non-laser and laser techniques with overall high efficacy and safety. Variation in outcomes between the two approaches does exist with limited comparative evidence in the literature. We sought to compare non-laser and laser TLE in a meta-analysis.

METHODS AND

RESULTS

We searched Medline, Embase, Scopus, ClinicalTrials.gov, and CENTRAL databases for TLE studies published between 1991 and 2021. From the included 68 studies, safety and efficacy data were carefully evaluated and extracted. Aggregated cases of outcomes were used to calculate odds ratio (OR), and pooled rates were synthesized from eligible studies to compare non-laser and laser techniques. Subgroup comparison of rotational tool and laser extraction was also performed. Non-laser in comparison with laser had lower procedural mortality (pooled rate 0% vs. 0.1%, P < 0.01), major complications (pooled rate 0.7% vs. 1.7%, P < 0.01), and superior vena cava (SVC) injury (pooled rate 0% vs. 0.5%, P < 0.001), with higher complete success (pooled rate 96.5% vs. 93.8%, P < 0.01). Non-laser comparatively to laser was more likely to achieve clinical [OR 2.16 (1.77-2.63), P < 0.01] and complete [OR 1.87 (1.69-2.08), P < 0.01] success, with a lower procedural mortality risk [OR 1.6 (1.02-2.5), P < 0.05]. In the subgroup analysis, rotational tool compared with laser achieved greater complete success (pooled rate 97.4% vs. 95%, P < 0.01) with lower SVC injury (pooled rate 0% vs. 0.7%, P < 0.01).

CONCLUSION

Non-laser TLE is associated with a better safety and efficacy profile when compared with laser methods. There is a greater risk of SVC injury associated with laser sheath extraction.

Для цитирования: Akhtar Zaki, Kontogiannis Christos, Georgiopoulos Georgios, Starck Christoph T, Leung Lisa W M, Lee Sun Y, Lee Byron K, Seshasai Sreenivasa R K, Sohal Manav, Gallagher Mark M. Comparison of non-laser and laser transvenous lead extraction: a systematic review and meta-analysis.. Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology. 2023. DOI: 10.1093/europace/euad316.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.