DOI 10.1093/europace/euad237

Медицинская библиотека · Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology · 2023

Mechanisms, predictors, and evolution of severe peri-device leaks with two different left atrial appendage occluders.

Lakkireddy Dhanunjaya, Nielsen-Kudsk Jens Erik, Windecker Stephan, Thaler David, Price Matthew J, Gambhir Alok, Gupta Nigel, Koulogiannis Konstantinos, Marcoff Leo, Mediratta Anuj, Anderson Jordan A, Gage Ryan, Ellis Christopher R

Рандомизированное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

AIMS: Incomplete left atrial appendage occlusion (LAAO) due to peri-device leak (PDL) is a limitation of the therapy. The Amulet IDE trial is the largest randomized head-to-head trial comparing the Amulet and Watchman 2.5 LAAO devices with fundamentally different designs. The predictors and mechanistic factors impacting differences in PDLs within the Amulet IDE trial are assessed in the current analysis.

METHODS AND

RESULTS

An independent core lab analysed all images for the presence or absence of severe PDL (>5 mm). The incidence, mechanistic factors, predictors using propensity score-matched controls, and evolution of severe PDLs through 18 months were assessed. Of the 1878 patients randomized in the trial, the Amulet occluder had significantly fewer severe PDLs than the Watchman device at 45 days (1.1 vs. 3.2%, P < 0.001) and 12 months (0.1 vs. 1.1%, P < 0.001). Off-axis deployment or missed lobes were leading mechanistic PDL factors in each device group. Larger left atrial appendage (LAA) dimensions including orifice diameter, landing zone diameter, and depth predicted severe PDL with the Watchman device, with no significant anatomical limitations noted with the Amulet occluder. Procedural and device implant predictors were found with the Amulet occluder attributed to the learning curve with the device. A majority of Watchman device severe PDLs did not resolve over time through 18 months.

CONCLUSION

The dual-occlusive Amplatzer Amulet LAA occluder provided improved LAA closure compared with the Watchman 2.5 device. Predictors and temporal observations of severe PDLs were identified in the Amulet IDE trial.

CLINICAL TRIAL REGISTRATION: https://clinicaltrials.gov Unique identifier NCT02879448.

Для цитирования: Lakkireddy Dhanunjaya, Nielsen-Kudsk Jens Erik, Windecker Stephan, Thaler David, Price Matthew J, Gambhir Alok, Gupta Nigel, Koulogiannis Konstantinos, Marcoff Leo, Mediratta Anuj, Anderson Jordan A, Gage Ryan, Ellis Christopher R. Mechanisms, predictors, and evolution of severe peri-device leaks with two different left atrial appendage occluders.. Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology. 2023. DOI: 10.1093/europace/euad237.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.