Virtually Wall-Less versus Standard Thin-Wall Venous Cannula in the Minimally Invasive Mitral Valve Surgery: Single-Center Experience.
Аннотация
Background and
Objectives
Minimally invasive cardiac surgery (MICS) has been developing since 1996. Peripheral cannulation is required to perform MICS, and good venous drainage and a bloodless field are crucial for the success of this procedure. We assessed the benefits of using a virtually wall-less cannula in comparison with the standard thin-wall cannula in clinical practice. Materials and
Methods
Between January 2021 and December 2022, we evaluated 65 elective patients, who underwent isolated minimally invasive mitral valve surgery. Both the virtually wall-less and the thin-wall cannulas were placed through a surgical cut-down. Patients' characteristics at baseline were similar in the two groups, except for the body surface area (BSA), which was greater in the virtually wall-less group compared to the thin-wall one. In the standard group, the size of the cannula was chosen depending on the patient's BSA, and the choice of the Smartcannula was based on their height.
Results
There were no significant differences between the two groups in terms of negative pressure applied, target flow achieved, hemolysis, the need for blood transfusion, and the post-operative increases in liver and renal enzymes. However, in all the patients, the estimated target flow was achieved, thereby showing the better hemodynamic performance of the virtually wall-less cannula, since, in this group, the patients' BSA was significantly greater compared to the thin-wall group. Ultimately, the mean cross-clamp time, as an indirect index of the effectiveness of the venous drainage, is shorter in the virtually wall-less group compared with the thin-wall group.
Conclusions
The virtually wall-less cannula should be preferred in minimally invasive mitral valve surgery due to its superior performance in terms of venous drainage compared with the standard thin-wall cannula.