DOI 10.1016/j.ejrad.2023.110722

Медицинская библиотека · European journal of radiology · 2023

Slice positioning in phase-contrast MRI impacts aortic stenosis assessment.

Troger Felix, Tiller Christina, Reindl Martin, Lechner Ivan, Holzknecht Magdalena, Pamminger Mathias, Poskaite Paulina, Kremser Christian, Ulmer Hanno, Gizewski Elke Ruth, Bauer Axel, Reinstadler Sebastian, Metzler Bernhard, Klug Gert, Mayr Agnes

СтатьяПеревод пока не подготовлен

Аннотация

AIMS: To determine the phase-contrast cardiovascular magnetic resonance imaging (PC-CMR) slice-position above aortic leaflet-attachment-plane (LAP) that provides flow-velocity, -volume and aortic valve area (AVA) measurements with best agreement to invasive and echocardiographic measurements in aortic stenosis (AS).

METHODS AND

RESULTS

Fifty-five patients with moderate/severe AS underwent cardiac catheterization, transthoracic echocardiography (TTE) and CMR. Overall, 171 image-planes parallel to LAP were measured via PC-CMR between 22 mm below and 24 mm above LAP. AVA via PC-CMR was calculated as flow-volume divided by peak-velocity during systole. Stroke volume (SV) and AVA were compared to volumetric SV and invasive AVA via the Gorlin-formula, respectively. Above LAP, SV by PC-CMR showed no significant dependence on image-plane-position and correlated strongly with volumetry (rho: 0.633, p < 0.001, marginal-mean-difference (MMD): 1 ml, 95 % confidence-interval (CI): -4 to 6). AVA assessed in image-planes 0-10 mm above LAP differed significantly from invasive measurement (MMD: -0.14 cm2, 95 %CI: 0.08-0.21). In contrast, AVA-values by PC-CMR measured 10-20 mm above LAP showed good agreement with invasive determination without significant MMD (0.003 cm2, 95 %CI: -0.09 to 0.09). Within these measurements, a plane 15 mm above LAP resulted in the lowest bias (MMD: 0.02 cm2, 95 %CI:-0.29 to 0.33). SV and AVA via TTE correlated moderately with volumetry (rho: 0.461, p < 0.001; bias: 15 ml, p < 0.001) and cardiac catheterization (rho: 0.486, p < 0.001, bias: -0.13 cm2, p < 0.001), respectively.

CONCLUSION

PC-CMR measurements at 0-10 mm above LAP should be avoided due to significant AVA-overestimation compared to invasive determination. AVA-assessment by PC-CMR between 10 and 20 mm above LAP did not differ from invasive measurements, with the lowest intermethodical bias measured 15 mm above LAP.

Для цитирования: Troger Felix, Tiller Christina, Reindl Martin, Lechner Ivan, Holzknecht Magdalena, Pamminger Mathias, Poskaite Paulina, Kremser Christian, Ulmer Hanno, Gizewski Elke Ruth, Bauer Axel, Reinstadler Sebastian, Metzler Bernhard, Klug Gert, Mayr Agnes. Slice positioning in phase-contrast MRI impacts aortic stenosis assessment.. European journal of radiology. 2023. DOI: 10.1016/j.ejrad.2023.110722.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.